DEVELOPMENT OF A GEOMETRIC CHARACTERIZATION METHODOLOGY FOR THE ASSESSMENT OF DUAL-MOBILITY LINERS USED IN TOTAL HIP ARTHROPLASTY

The British Orthopaedic Research Society (BORS) Annual Meeting 2021, held online, 13–14 September 2021.



Abstract

Abstract

OBJECTIVES

Dual mobility (DM) total hip replacements (THRs) were introduced to reduce the risk of hip dislocation in at-risk patients. DM THRs have shown good overall survivorship and low rates of dislocation, however, the mechanisms which describe how these bearings function in-vivo are not fully understood. This is partly due to a lack of suitable characterisation methodologies which are appropriate for the novel geometry and function of DM polyethylene liners, whereby both surfaces are subject to articulation. This study aimed to develop a novel semi-quantitative geometric characterisation methodology to assess the wear/deformation of DM liners.

METHODS

Three-dimensional coordinate data of the internal and external surfaces of 14 in-vitro tested DM liners was collected using a Legex 322 coordinate measuring machine. Data was input into a custom Matlab script, whereby the unworn reference geometry was determined using a sphere fitting algorithm. The analysis method determined the geometric variance of each point from the reference surface and produced surface deviation heatmaps to visualise areas of wear/deformation. Repeatability of the method was also assessed.

RESULTS

Semi-quantitative analysis of the surface deviation heatmaps revealed circumferential damage patterns similar to those reported in the literature. Additionally, the location of the damaged regions corresponded between the internal and external surfaces. Comparing five repeat measurements of the same liner, the maximum geometric variance of each surface varied by 1 µm (standard deviation) suggesting a high repeatability of the method.

CONCLUSIONS

This study presents an effective and highly repeatable characterisation methodology to semi-quantitatively assess the wear/deformation of in-vitro tested DM liners. This method is suitable for the analysis of retrieved DM liners whereby no pre-service information is available, which may provide information about the complex in-vivo kinematics and mechanical failure mechanisms of these bearings.