CONVERTIBLE TOTAL ELBOW PROSTHESIS: IN VITRO ASSESSMENT OF THE STRENGTH OF THE AXLE MODULAR CONNECTION UNDER LOW AND HIGH DEMAND BIOMECHANICAL CONDITIONS IN A NOVEL DEVICE IN LINKED CONFIGURATION

International Society for Technology in Arthroplasty (ISTA) meeting, 32nd Annual Congress, Toronto, Canada, October 2019. Part 2 of 2.



Abstract

Introduction

Total Elbow Arthroplasty (TEA) is recognized as an effective treatment solution for patients with rheumatoid arthritis or for traumatic conditions. Current total elbow devices can be divided into linked or unlinked design. The first design usually presents a linking element (i.e. an axle) to link together the ulnar and humeral components to stabilize the joint; the second one does not present any linkage and the stability is provided by both intrinsic design constraints and the soft tissues. Convertible modular solutions allow for an intraoperative decision to link or unlink the prosthesis; the modular connections introduce however additional risks in terms of both mechanical strength and potential fatigue and fretting phenomena that may arise not only due to low demand activities loads, but also high demand (HD) ones that could be even more detrimental. The aim of this study was to assess the strength of the modular connection between the axle and the ulnar component in a novel convertible elbow prosthesis design under simulated HD and activities of daily living (ADLs) loading.

Methods

A novel convertible total elbow prosthesis (LimaCorporate, IT) comprising both ulnar and humeral components that can be linked together by means of an axle, was used. Both typical ADLs and HD torques to be applied to the axle were determined based on finite element analysis (FEA); the boundary load conditions for the FEA were determined based on kinematics analysis on real patients in previous studies. The FEA resultant moment acting on the axle junction during typical ADLs (i.e. feeding with 7.2lbs weight in hand) was 3.2Nm while for HD loads (i.e. sit to stand) was 5.7 Nm. In the experimental setup, 5 axle specimens coupled with 5 ulnar bodies through a tapered connection (5 Nm assembly torque) were fixed to a torque actuator (MTS Bionix) and submerged in a saline solution (9g/l). A moment of 3.2 Nm was applied to the axle for 5M cycles through a fixture to test it under ADLs loading. After 5M cycles, the axles were analyzed with regards to fretting behavior and then re-assembled to test them against HD loading by applying 5.7 Nm for 200K cycles (corresponding to 20 years function).

Results

All 5 samples withstood all 5.2M loading cycles without any mechanical failure. At the end of 5M cycles, each axle was still stable as the measured disassembly torque was 3.96 +/−0.18 Nm. Slight signs of fretting were detected on the tapered connection after 5M cycles, however they did not compromise the mechanical connection nor the stability.

Discussion and Conclusions

Currently there are no reference standards that properly define protocols for biomechanical testing of elbow prostheses. In the present study, a test to mechanically assess the strength of an axle connection under both typical ADLs and HD loads was set. The connection was able to withstand the imposed conditions. In general, testing of TEA devices should include not only standard ADLs loads but also HD loads, which could be more detrimental for the long-term survivorship.