Abstract
A pre-clinical experimental simulation model has been previously successfully developed, and was shown to have the potential for investigation of the biomechanical and tribological performance of early stage knee therapies. In order to investigate interventions that may necessitate sacrifice of the natural ligaments, it is necessary to replicate their function. This study investigated the most effective spring constraint conditions for the porcine knee model with the aim of replicating the natural ligament function.
The replication of natural ligament function was achieved through the use of physical springs in the anterior-posterior (AP) axis. Spring-9 (9 N/mm) and spring-20 (20 N/mm) were set at different free lengths in a natural knee simulator. The A/P displacement and shear force outputs from porcine knee samples (N=6) were measured and the most appropriate spring setting was determined by comparing the outputs at different spring settings with intact knee.
The A/P displacement of both spring-9 and spring-20 showed similar shapes to the all ligament control. Spring-9 with a free length of 4 mm and spring-20 with a free length of 5 mm showed minimal differences in A/P displacement output compared to the all ligament controls. There was no statistical difference between the two minimal differences either in A/P displacement or in shear force (paired t-test, p>0.05), which indicated that both conditions were appropriate spring constraint settings for the natural porcine knee model.
A porcine knee simulation model with refined spring constraint conditions was successfully developed in this study. Human knee model is currently under investigation using the methodology developed in porcine knee model, which will be more appropriate to investigate the effect of early stage knee therapies on the tribological function of the natural knee.