BIOMECHANICAL COMPARISON OF AUGMENTED LOCKING PLATE FIXATION VERSUS CONVENTIONAL LOCKING PLATING

The European Orthopaedic Research Society (EORS) 25th Annual and Anniversary Meeting, Munich, Germany, September 2017. Part 2 of 2.



Abstract

Introduction

The incidence of distal femoral fractures in the geriatric population is growing and represents the second most common insufficiency fracture of the femur following fractures around the hip joint. Fixation of fractures in patients with poor bone stock and early mobilisation in feeble and polymorbide patients is challenging. Development of a fixation approach for augmentation of conventional LISS (less invasive stabilization system) plating may result in superior long-term clinical outcomes and enhance safe weight bearing.

Objectives

The aim of this study was to investigate the biomechanical competence of two different techniques of augmented LISS plating for treatment of osteoporotic fractures of the distal femur in comparison to conventional LISS plating.

Materials & methods

Unstable distal femoral fracture AO/OTA 33-A3 was set in artificial femora with low density simulating osteoporotic bone. Three study groups, consisting of 10 specimens each, were created for instrumentation with a 9-hole LISS plate, a LISS plate with an additional 3D-printed polyactide cylindrical intramedullary graft, as well as a LISS plate plus a medial 3.5mm LCP (locking compression plate) - double plating. All specimens were non-destructively tested under axial (20–150N) and torsional (0–4Nm) quasi-static loading. Each construct was tested with two different working length (WL) configurations (long and short) of the LISS plate. Relative movements between the most medial superior and inferior osteotomy aspects were investigated via three-dimensional motion tracking analysis.

Results

Interfragmentary displacement along the femur axis (mm) under 150N axial loading was 2.03±0.23/1.65±0.27 for LISS with long/short WL, 0.18±0.06/0.18±0.04 for double plating with long/short WL, and 0.40±0.05/0.30±0.05 for LISS plus graft with long/short WL. Shear interfragmentary displacement (mm) under 4Nm torsional loading in internal rotation was 1.16±0.17/0.92±0.11 for LISS with long/short WL, 0.40±0.10/0.43±0.07 for double plating with long/short WL, and 1.09±0.13/0.82±0.11 for LISS plus graft with long/short WL.

Double plating revealed significantly smaller longitudinal and shear displacement compared to the other two techniques for long and short WL, respectively (P≤0.010). In addition, LISS plus graft fixation was with significantly less longitudinal displacement in comparison to conventional LISS plating for long and short WL, respectively (P≤0.001).

Long WL resulted in significantly higher longitudinal and shear displacement compared to short WL for LISS and LISS plus graft (P≤0.032), but not for double plating (P=1.000).

Conclusion

Intramedullary grafting resulted in significantly increased fracture stability under axial loading in comparison to conventional LISS plating.

However, it was not efficient enough to achieve comparable stability to double plating.