Abstract
Introduction
Acetabular component positioning, offset, combined anteversion, leg length, and soft tissue envelope around the hip plays an important role in hip function and durability. In this paper we will focus on acetabular positioning of the cup.
Technique
The axis of the pelvis is identified intra-operatively as a line drawn from the highest point of the iliac crest to the middle of the greater trochanter. Prior to reaming the acetabulum, an undersized trial acetabular component is placed parallel and inside the transverse ligament, inside the anterior column and projecting posterior to the axis of the pelvis. This direction is marked and the subsequent reaming and final component placement is performed in the same direction. The lateral opening is judged based on the 45-degree angle from the tear drop to the lateral margin of the acetabulum on anteroposterior pelvic radiographs. The final anteversion of the cup is adjusted based on increased or decreased lumbar lordosis and combined anteversion.
Methods
Anteroposterior pelvic radiographs of 100 consecutive patients undergoing posterior THR between September 2010 and March 2011 with this method were evaluated for cup inclination angle and anteversion using EBRA software.
Results
There were no malalignments or dislocations. The mean cup inclination angle and anteversion were 41 ± 5.1 degrees (range 37.1 – 48.4) and 22.1 ± 4.8 degrees (range 16.6 – 29.3), respectively.
Conclusion
This is a reproducible method of cup positioning and with proper femoral component position and restoring leg length, offset, combined anteversion, and balance soft tissue around the hip. These factors affect the incidence of dislocation, infection, reduced wear, and durability.