header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Trauma

A REPLACEMENT FOR THE KIRSCHNER WIRE: THE POTENTIAL OF A BIOABSORBABLE PIN IN PAEDIATRIC FRACTURE MANAGEMENT

Scottish Committee for Orthopaedics and Trauma (SCOT), September 2017 Meeting



Abstract

Kirschner wires are commonly used in paediatric fractures, however, the requirement for removal and the possibility of pin site infection provides opportunity for the development of new techniques that eliminate these drawbacks. Bioabsorbable pins that remain in situ and allow definitive closure of skin at the time of insertion could provide such advantages.

Three concurrent studies were performed to assess the viability of bioabsorbable pins across the growth plate. (1) An epidemiological study to identify Kirschner wire infection rates. (2) A mechanical assessment of a bioabsorbable pin compared to Kirschner wires in a simulated supracondylar fracture. (3) The insertion of the implants across the physis of sheep to assess effects of the bioabsorbable implant on the growth plate via macroscopic, pathohistological and micro-CT analysis.

An infection rate of 8.4% was found, with a deep infection rate of 0.4%. Mechanically the pins demonstrated comparable resistance to extension forces (p=) but slightly inferior resistance to rotation (p=). The in vivo component showed that at 6 months: there was no leg length discrepancy (p=0.6), with micro-CT evidence of normal physeal growth without tethering, and comparable physeal width (p=0.3).

These studies combine to suggest that bioabsorbable pins do not represent a threat to the growth plate and may be considered for physeal fracture fixation.


Email: