258. CONSEQUENCES OF EDGE LOADING ON THE FRICTION COEFFICIENT OF CERAMIC-ON-CERAMIC TOTAL HIP ARTHROPLASTY: APPLICATION FOR SQUEAKING



Abstract

Purpose of the study: Ceramic-on-ceramic THA explants exhibit a higher wear rate than that predicted by classical simulators. This appears to be related to edge loading, which could perhaps be reproducible in vitro by creating a microseparation between the two components. The purpose of this study was to evaluate this coefficient of friction for ceramic-on-ceramic THA with edge loading. This should enable prediction of wear in the event of microseparation.

Material and methods: Three 32mm alumina inserts (Biolox Forte Ceramtec®) were tested on a friction simulatior (Prosim®). The cup was positioned vertically (75° inclination) to reproduce edge loading. The metal-back and the acetabular insert were sectioned to avoid impingement between the neck and cup. Contact was imposed along the border of the cup, then perpendicularly to it. The tests were performed under lubrication conditions (25% bovine serum). In order to simulate severe contact pressures, the tests were also conducted with a third body inserted between the head and the edge of the cup. To obtain reference values of the centred regimen, tests were first run with identical components positioned horizontally.

Results: Edge loading was achieved for cups inclined at 75°. The coefficient of friction was 0.02±0.001 under centred conditions. For edge loading conditions, the coefficient of friction was significantly increased, to a mean 0.09±0.00 for movement along the acetabular border and 0.034±0.001 for movement perpendicular to the border. Squeaking occurred for 15 s when the third body was introduced, corresponding to a coefficient of friction 15-fold higher (0.32±0.003) than under ideal conditions.

Discussion: For the first time, the coefficient of friction of edge loading was determined under conditions of lubrication. The friction coefficient of ceramic-on-ceramic THA was greater for a very vertical cup, but remained (0.1) equivalent to the metal-on-metal coefficient under optimal conditions. When a third body was introduced, transient squeaking occurred with a very high coefficient of friction.

Conclusion: Implantation of cups with a high abduction angle induces edge loading and an increased coefficient of friction, and should be avoided.