Abstract
Purpose of the study: The purpose of this study was to use weight-bearing radiographies to study the mobility of the polyethylene insert in relation to the femoral and tibial components of a total knee arthroplasty (TKA) with preservation of the posterior cruciate ligament and a mobile plateau with rotation and anterioposterior translation (INNEX® Anterior-Posterior Glide, Zimmer).
Material and methods: A 3D kinematic study of the femoral and tibial component and the mobile insert was conducted on a series of 51 first-intention TKA using a computer-assisted matching system between 3D prosthetic models and the radiographic silhouette of the implants.
Results: At mean 23 months postoperatively, the poly-ethylene tibial insert exhibited an increase in its internal rotation during flexion. This rotation, knee extended, was limited to rotation between the insert and the tibial base. With increased flexion, there was an increase in the value and the portion of rotation involving the femoral component in relation with the mobile tibial insert.
Discussion: The degree of insert mobility has varied depending on the report. Certain authors have reported relatively limited mobility because of a minimally congruent superior surface allowing anteroposterior and mediolateral translation as the femur glided over the insert. Others report mobility of the mobile plateau in relation to the tibial base and minimal rotation of the femoral component. Rotation of the polyethylene insert in TKAs with a mobile plateau appears to be quite variable. With the LCS AP Glide prosthesis, anteroposterior translation of the mobile plateau was measured at a mean 5.6 mm (1–1.125 mm). Paradoxical anterior translation, rather than posterior translation, of the mobile plateau with flexion has been reported in a few patients.
Conclusion: The mobile plateau has exhibited progressive increase in internal rotation with flexion. We have concluded that the major part of the mobility occurs between the mobile plateau and the tibial base. However, with flexion, the femoral component increased its mobility over the plateau. During flexion, anteroposterior translation occurred between the femoral piece and the tibial insert, and between the tibial insert and the tibial base, but the direction of the translation of the mobile tibial insert appeared to be unpredictable with the non-constrained prosthesis used for this study.