Abstract
Most of computer-assisted computer assisted system rely on the peri-operative acquisition of the anterior pelvic plane defined as the plane crossing the two anterior iliac spine and the symphysis. The goal of this study was to evaluate in vivo and in vitro the accuracy of the anterior pelvic plane acquisition, considered as the reference for computer-assisted total hip arthroplasty (THA).
Cup placement was performed using an imageless computer-assisted system in thirty patients during THA. Post-operatively the position of the cup was evaluated on computed tomography using a validated tridimensional software. The differences between the perioperative and postoperative angles for abduction and anteversion were compared using a two-group pair test.
On two cadavers four clinicians performed ten times the anterior pelvic plane acquisition using three Methods: percutaneously, with ultrasound and by direct bony acquisition defined as the reference. The mean error for each anterior pelvic plane acquisition method was compared using a univariate variance model for repeated measurements.
In vivo, the mean difference between the perioperative and postoperative abduction angles was 4° and not statistically significant. For anteversion, the difference was 4° and not significant in patients with BMI < 27. The difference was 11° and significant in patients with BMI > 27 (p< 0.001). In vitro, the mean errors for rotation and tilt were respectively 3.8 ° and 19.25 ° for cutaneous acquisition, 2.8° and 6.2° for ultrasound acquisition method. The errors were statistically higher with the percutaneous method (p< 0.001).
According to our results, the accuracy of the standard percutaneous acquisition method of the anterior pelvic plane in computer-assisted THA is limited. The ultrasound acquisition method may represent a reliable alternative.