Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

A1212. METAL ON METAL HIP ARTHROPLASTY: WHERE ARE WE IN 2009?



Abstract

Metal-on-metal bearings have become popular in the last ten years because of a low wear rate combined with the ability to use large head sizes for conventional total hip arthroplasty (THA) and to facilitate resurfacing hip arthroplasty. Further advantages of metal-metal bearings include the fact that they are not at risk for fracture, and they can be made as modular or non-modular acetabular implants.

It was recognized early that metal-on-metal implants had the potential to increase serum ion levels, and this was demonstrated in a number of studies. The significance of elevated ion levels, however, for most patients has been primarily a theoretical concern of toxicity, carcinogenesis or mutagenicity, and to date very few, if any, systemic problems related to systemic metal ions have been documented with certainty. Nevertheless, most surgeons have avoided use of the implants in patients who are likely to become pregnant, patients with renal disease, or patients with major systemic illnesses which have a high likelihood of leading to renal disease. Furthermore, most have avoided using them in patients with known dermal metal allergies, even though the connection between dermal metal allergies and metal bearings has not been established.

Unexpectedly, an extremely important concern has emerged with metal bearings: the finding of local inflammatory reactions related to metal bearings. These inflammatory reactions can take several forms including pain with a milky effusion, local tissue necrosis, or large fluid collections or pseudotumors. The histology of these different reactions appears to be predominantly lymphocytic in nature and a term for at least some of these reactions has been coined “AVALS”. Whether these local reactions are primarily immunologic in nature or primarily related to dose of local metal ions or debris remains uncertain. While there is much still to be learned, it appears that certain patient populations may be at increased risk for metal reactions, possibly related to implant size (women and smaller patients). It also seems verticallyoriented implants, which create edge loading, increase wear and increase risk of local metal reactions.

Perhaps the most important question is the incidence of local metal reactions, which remains to be defined. To date the problems in most series have been infrequent, less than 1 or 2 percent. However, in a few selected series the incidence has been higher, and when screening has been done for asymptomatic patients with fluid or masses around the joint, the rate has been higher in at least one reported series.

Surgeons may interpret the importance of local metal reactions differently, but certainly ultimately incidence of this problem will have a very major effect on the future of these bearings.

Correspondence should be addressed to Diane Przepiorski at ISTA, PO Box 6564, Auburn, CA 95604, USA. Phone: +1 916-454-9884; Fax: +1 916-454-9882; E-mail: ista@pacbell.net