Abstract
Many types of bioactive bone cement have been developed to overcome the disadvantages of polymethyl-methacrylate (PMMA) bone cement, especially its lack of bone-bonding ability, which occasionally leads to aseptic loosening of prostheses used for arthroplasty. Earlier, we showed that bioactive bone cements containing either nano-sized or micron-sized titania (TiO2) particles had excellent in vivo osteoconductivity.
However, anatase phase titania particles contained in these bioactive bone cements raise concerns about their safety in vivo. We developed pure rutile micron-sized titania particles. because rutile is the only stable phase, whereas anatase is metastable.
In this study, polymethylmethacrylate (PMMA)-based bone cement containing pure rutile micron-sized titania (TiO2) particles were developed, and their mechanical properties and osteoconductivity are evaluated. The three types of bioactive bone cement were T10, T20, and T30, which contained 10, 20, and 30wt% TiO2, respectively.
Commercially available PMMA cement (PMMAc) was used as a control. Hardened cylindrical cement sample (φ2.5mm*10mm) was inserted manually on rabbit femur vertically. Push out test was performed for evaluation of bonding strength. For mechanical testing, the flexural strength, flexural modulus, and compressive strength were measured.
Results of this study revealed that polymethylmeth-acrylate (PMMA)-based bone cement containing pure rutile micron-sized titania particles has outstanding osteoconductivity in vivo, and their mechanical properties were exceeded that of commercially available PMMA cement. Interfacial shear strength of T10, T20 and T30 were 17.1~24.0MPa each at 12 weeks, and were significantly higher than PMMAc. In general, the interfacial bonding strength of bone cement depends mainly on its interdigitation with cancellous tissue, which is accomplished by the pressurized injection of the cement in paste form. On the other hand, we inserted the hardened specimens into oversized holes on rabbit femur in this study, because we intended to examine the osteoconductive and bone-bonding potentials of each material. The flexural strength, flexural modulus, and compressive strength were equivalent to or exceeded that of PMMAc.
These results show that bone cement containing pure rutile micron-sized titania particles is a promising material applied to prosthesis fixation as well as vertebroplasty.
Correspondence should be addressed to Diane Przepiorski at ISTA, PO Box 6564, Auburn, CA 95604, USA. Phone: +1 916-454-9884; Fax: +1 916-454-9882; E-mail: ista@pacbell.net