Full Access

A PROSPECTIVE RANDOMISED TRIAL COMPARING THE LONG GAMMA NAIL WITH THE SLIDING HIP SCREW FOR THE TREATMENT OF UNSTABLE PERTROCHANTERIC HIP FRACTURES



Abstract

Controversy exists whether to treat unstable pertrochanteric hip fractures with either intramedullary or extramedullary devices. A prospective randomised control trial was performed to compare the outcome of unstable pertrochanteric hip fractures stabilised with either a sliding hip screw (SHS) or Long Gamma Nail (LGN). The hypothesis was that there is no difference in outcome between the two modes of treatment.

Over a four year period, 210 patients presenting with an unstable pertrochanteric hip fracture (AO/OTA 31 A2.1/A2.2/A2.3) were recruited into the study. Eligible patients were randomised on admission to either LGN or SHS. Follow-up was arranged for three, six, and twelve months. Primary outcome measures were implant failure and implant ‘cut-out’. Secondary measures included mortality, length of hospital stay, and EuroQol outcome score.

Five patients required revision surgery for implant cutout, of which three were LGNs and two were SHSs (no significant difference). There was a significant correlation between tip apex distance and the need for revision surgery. There were no incidences of implant failure or deep infection. Mortality rates between the two groups were similar when corrected for mini mental score. There was no difference between the two groups with respect to tip apex distance, hospital length of stay, blood transfusion requirement, and EuroQol outcome score.

The sliding hip screw remains the gold standard in the treatment of unstable pertrochanteric fractures of the proximal femur.

Correspondence should be addressed to Editorial Secretary Mr ML Costa or Assistant Editorial Secretary Mr B.J. Ollivere at BOA, 35–43 Lincoln’s Inn Fields, London WC2A 3PE, England; Email: mattcosta@hotmail.com or ben@ollivere.co.uk