7.P.02 EFFECTIVE ANTIMIROBIAL ACTIVITY OF SILVER-COATINGS AGAINST SMALL COLONY VARIANTS



Abstract

Megaendoprotheses are widely used in the reconstruction of large bone defects in orthopaedic tumour surgery. The major complications (up to 36%) are periprosthetic infections. Persisting periprosthetic infections lead to secondary amputation up to 37% of the cases. One underestimated reason for persisting infections are subpopulations of S. aureus called “small colony variants” (SCVs). Aim of this study was to evaluate that silver ions might prevent or cure a periprosthetic infection caused by SCVs.

For testing the antimicrobial activity of silver-coated titanium we used a technique introduced by Bechert et al. Therefore an adhesion and proliferation assay was performed with clinical isolates of S. aureus SCV (A22616/3). We tested the adhesion and proliferation properties of S aureus SCV on stainless steel (steel), Cobalt-Chrome-Molybdenum-alloy (CoCrMo), Titan-Aluminium-Vanadium-alloy (TiAlVa) and silver-coated Titan-Aluminium-Vanadium-alloy (scTiAlVa).

Adhesion of S. aureus SCV is significantly reduced on scTiAlVa vs. steel (p> 0001). We could also demonstrate that the proliferation rate of scTiAlVa vs. all tested materials is significant (p> 0001) lower.

We concluded that silver-coating has an effective antimicrobial activity against S. aureus SCVs. Thus silver-coated megaendoprostheses are a good prophylaxis against persisting infections caused by S. aureus SCVs.