Abstract
A revision of a first generation KMFTR prosthesis due to stem breakage is a problem oncologic surgeons are regularly faced with. We designed an adapter which allows us to connect new MUTARS components to the original KMFTR devices. Thus it is possible to bypass an exchange of the whole prosthesis.
We used this adapter in 10 patients. Time of revision was in average 16.6 years after primary implantation of the KMFTR prostheses. Reasons for revision were femoral stem breakage (n = 5), breakage of the tibial component (n = 3) and periprosthetic fracture (n = 2, one femoral, one tibial).
The femoral stem (3 cases) and the tibial stem (2 cases) as well as the tibial plateau and body (2 cases) could be replaced by MUTARS parts and conjoined with the remaining KMFTR devices. Three cases were converted to MUTARS total femur.
Postoperative complications were one aseptic loosening and one cone-dislocation. Pre-incidence function was restored in all cases. The average Musculoskeletal-Tumour-Society-Score was 82.9% of normal function.
The results show that the new adapter facilitates to restore pre-incidental extremity function by performing a relatively modest revision.
Correspondence should be addressed to Professor Stefan Bielack, Olgahospital, Klinikum Stuttgart, Bismarkstrasse 8, D-70176 Stuttgart, Germany. Email: s.bielack@klinikum_stuttgart.de