C-STEM VERSUS CHARNLEY. RESULTS FROM A RANDOMIZED TRIAL AFTER TWO YEARS WITH RADIOSTEREOMETRY (RSA)



Abstract

Introduction: C-stem is a triple taper polished femoral stem. The rationale for this design is to achieve an evenly distributed proximal loading of the cement mantle. This design is thought to enhance stability of the stem inside the mantle and lead to bone remodelling medially. There is to our knowledge no randomized trial comparing this stem to a well documented stem. We chose to compare it to the best documented stem in the Norwegian arthroplasty register, the Charnley monoblock.

Methods: 70 patients scheduled for total hip replacement were randomized to either C-stem or Charnley monoblock. All received a 22 mm stainless steel head, OGGEE cup and Palacos Cement with Gentamycin. We used a transgluteal approach in all cases. Harris and Oxford hip scores were measured preoperatively and after two years. Standard X-rays were taken postoperatively and after two years. Radiostereometry (RSA) was done postoperatively and after 3,6,12 and 24 months.

Results: There was no significant difference in Harris or Oxford hip scores after two years.

RSA after two years: (table deleted)

Discussion: Polished tapered stems are designed to sink inside the mantle. Our results confirm this theory for the C-stem. The subsidence is comparable to other collarless tapered stems with good long-term survival. For all other migrations/rotations the C-stem is as stable as the Charnley monoblock. This predicts good long-term results for this stem.

Correspondence should be addressed to BHS c/o BOA, at the Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London, WC2A 3PE, England.