Abstract
Introduction: Patient selection has been critical to the excellent medium-term clinical results following hip resurfacing. Hypersensivity to metal ion debris has been described in previous generations of metal-on-metal bearings. This may also be a problem that affects modern designs. Characteristic histological changes have been identified (ALVAL). There are few studies that include large female numbers, and show separate outcome and implant survival.
Methods: Eighty-one female hips at a mean of 23 months (11–43) had an ASR hip resurfacing procedure at a single-surgeon independent centre. Mean age was 55 years (28–69). Harris Hip Scores (HHS) were recorded at one-year follow-up. Failures were analysed.
Results: HHS improved from 46.4 (11–77) to 90.2 (27–100). Overall, there was a 7.4 % revision rate. There were 3 femoral neck fractures. In the entire series of 98 female patients there were 3 cases of severe pain requiring revision (3.1%). All three patients had HHS < 50 at one year follow-up. Patients had groin pain, reduced flexion and a painful straight leg raise. Blood results were not suggestive of infection. Aspiration of the hip joint in each case revealed copious amounts of milky green grey aseptic fluid. All had similar macroscopic changes at revision. There were characteristic histological changes in keeping with ALVAL. All 3 patients were revised to THRs with ceramic bearings.
Discussion: The failure rate of 7.4% in the older female group is poor at this early stage following hip resurfacing. The incidence of metal hypersensitivity in our series suggests this complication may be more common than previously thought. Patients with persisting pain of unknown aetiology following resurfacing at other centres may have metal hypersensitivity.
Correspondence should be addressed to Mr Peter Howard, Editorial Secretary, BHS, c/o BOA, The Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PE, England.