Studies suggest that specialty hospitals and high surgical volume decrease adverse outcomes related to hip arthroplasty. Little is known, however, concerning the influence of imageless computer navigation systems on a surgeon’s experience and subsequent placement of implants in the setting of hip resurfacing arthroplasty.
A retrospective review of 71 consecutive hip resurfacing arthroplasties placed with computer assisted navigation during 2006 and 2007 was performed. Forty-seven operative days encompassing the surgeon’s entire experience with hip resurfacing were analysed. Within this single surgeon series, operative time, intraoperative cup inclination and femoral stem/shaft angles, as well as postoperative cup inclination and femoral stem/shaft angles were measured and compared over three discreet, sequential operative time intervals.
Intraoperative cup inclination angles were comparable to postoperative radiographic values as there was no significant difference (p=.059). Computer assisted navigation produced consistent values despite different levels of surgeon experience in the setting of intraoperative cup inclination (42.8°, 43.5°, and 40.1°) and postoperative cup (46.1°, 43.9°, and 42.9°) and femoral stem (147.9°, 146.5°, and 144.0°) radiographic alignment. A statistically significant difference existed between intraoperative femoral stem/shaft angles compared to postoperative radiographs measurements (p< .001), however, all means maintained a valgus orientation compared to the native neck angle. There was a correlation between evolving surgeon experience and intraoperative stem placement (143.5°, 142.1°, and 138.0°, respectively) despite the mean values remaining well clustered (p< .001). Operative times significantly decreased (p< .001) with surgeon experience, showing the largest decrease after the 1st sequence interval (109.6, 97.8, and 94.8 min, respectively). No femoral notching (0/71) occurred throughout the series.
Computer assisted navigation provides a dependable method of accurate hip resurfacing arthroplasty component positioning as measured by cup inclination, in addition to a reliable technique for valgus stem placement and avoidance of notching. Furthermore, computer navigation allows for consistency and offers a protective effect on component alignment independent of surgeon procedural experience.
Correspondence should be addressed to Mr K Deep, General Secretary CAOS UK, Dept of Orthopaedics, Golden Jubilee National Hospital, Glasgow G81 4HX, Scotland. Email: caosuk@gmail.com