Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

THE EFFECT OF DISTRACTION VECTOR ORIENTATION ON THE REGENERATE TISSUE DURING BILATERAL MANDIBULAR DISTRACTION



Abstract

Bilateral mandibular lengthening is widely accepted during mandibular distraction osteogenesis. However, distraction osteogenesis are sometimes associated with clinical complications such as open bite deformity, lateral displacement of temporo-mandibular joint, premature consolidation and pin loosening. Although distraction osteogenesis aims to develop pure tensile strain on the regenerate tissue however, in real life situation due to differences in device orientation, materials and misalignment it is often subjected to complex stress and strain regimes.

The objective of this study was to characterise the mechanical environment (stress and strain) in the Finite Element Models (FEM) of regenerate tissue during two different device orientations:

  1. (a) device placed parallel to the mandibular body

  2. (b) device placed parallel to the axis of distraction.

Furthermore, the influence of misalignment from above two idealised orientations was also investigated.

The distraction protocol in this study was similar to previous study of Loboa et al (2005). FE models were developed at four stages: end of latency, distraction day two, distraction day five and distraction day eight. At each time period a distraction of 1mm was applied to the model as it is most widely used distraction rate. In these models two primary distraction vectors were simulated; first when the device is parallel to the body of the mandible and second when the device is parallel to the axis of distraction.

Results indicate that when the device is placed parallel to the mandible the effect of distraction vector variation due to misalignment in transverse plane (±150 at an interval of 50 ; + indicate lateral and indicates medial) is symmetric and variation in the stress and strain regimes on regenerative tissue are less than 3%. However, when the device is placed parallel to axis of distraction the corresponding change is asymmetric and almost double in magnitude. The greatest differences were seen when misalignment is towards lateral side (+150). Similarly in the sagittal plane variations up to 17% were developed due to 0- 400 change in the distraction vector orientation. Thus the orientation of device which determines the distraction vector plays an important role in determining the mechanical environment around regenerative tissue. The results suggest that implications of misalignment of the device and its sensitivity from the ideal situation should be well understood during clinical planning.

Correspondence should be addressed to David Haynes, PhD, Senior Lecturer, President ANZORS, at Discipline of Pathology, School of Medical Sciences, University of Adelaide, SA, 5005, Australia