Abstract
Large diameter metal-on-metal (MOM) bearings are becoming increasingly popular for young, active patients. Clearance is a particularly important consideration for designing MOM implants, considering historical experience of equatorial contact and high frictional torque. Lubrication theory predicts increasing the clearance will result in diminished lubrication, resulting in increased friction and wear. Clinical cases of transient squeaking in patients with resurfacing bearings have been noted in recent years, with some reporting an incidence of up to 10% between 6 months and 2 years post-implantation. This study aimed to investigate the impact of increasing clearance on the lubrication, friction and squeaking of a large diameter metal-on-metal resurfacing bearing through frictional studies.
Clinical-grade MOM implants of 55mm diameter and 100μm diametric clearance, and custom-made, 55mm bearings with diametric clearances of approximately 50μm and 200μm (DePuy International Ltd) were tested in a friction simulator. Components were inverted with a flexion-extension of ±25o applied to the head and lubricated with 25% and 100% newborn bovine serum. A peak load of 2kN, with swing-phase loads of 25N, 100N and 300N were applied.
Sound data was recorded during each friction test using a MP3 recorder and pre-amplifier. A microphone was set up at a distance of 50mm from the implant, and data recorded over a minimum of 10 seconds where sound was generated. Sound data was assessed through narrow band analysis on Frequency Master software (Cirrus Research, UK).
Lubrication was assessed by directly measuring the separation between the head and cup during the test cycle by ultrasonic methods (Tribosonics, UK). An ultrasound sensor was bonded to the back of the cup and reflection measurements were taken during the friction tests with a sampling rate of 100Hz. Using equations which related reflection coefficient to lubricant properties and thickness, values for the film thickness were calculated.
The surface replacement with the largest clearance yielded the highest friction factor for each test condition. The difference between the large clearance bearing and the smaller clearance samples was statistically significant in 25% bovine serum, the more clinically relevant lubricant (ANOVA, p< 0.05). The 50μm clearance group yielded similar results to the 100μm clearance bearing, although a slight increase in friction was observed.
Squeaking occurred during every test in the large clearance group. There was a reduced incidence of squeaking in the smaller clearances, with the lowest incidence observed in the 100μm clearance group.
The smallest separation of the head and cup was observed within the large clearance bearings. The best lubrication condition measured ultrasonically was observed within the 100μm clearance bearing. There appeared to be good correlation between friction, lubrication and the incidence of squeaking. This study suggests a large diametric clearance results in reduced lubrication, increased friction and an increased incidence of squeaking. However, there is a minimum diametric clearance that can be tolerated, as clearance must accommodate the manufacturing tolerance.
Correspondence should be addressed to Mr Carlos A. Wigderowitz, Senior Lecturer, University Dept of Orthopaedic and Trauma Surgery, Ninewells Hospital and Medical School, Dundee DD1 9SY