Abstract
This is a clinically based study to assess the reliability of fracture stiffness as a measurement of clinical union and investigate other indicators which may aid the clinician to accurately determine when fracture fixation may be removed.
A fracture bending stiffness in the sagittal plane of 15Nm/deg. has been stated as a satisfactory endpoint at which an external fixator may be removed from diaphyseal fractures of the tibia. However using this as a measure to determine when to remove support in a study of 76 patients 4 continued to a malunion. Fracture callus properties were measured in clinic. The fixator was removed for the tests and a specially designed system was used to measure displacement and load. Fracture stiffness was measured in different planes and at various loading rates. Passive stressing of the leg was performed whilst fracture displacement was recorded. A constant load was applied for a longer period to assess creep properties.
Fracture stiffness was found to vary between different planes of measurement and on load rate. The visco-elastic characteristics of the callus changed with time. In early measurements, the callus absorbed a large proportion of energy when a load was applied. Later tests showed a progressive change with the callus absorbing less energy. This demonstrates that the properties of the callus changed with time, with the viscous element diminishing and the elastic element increasing. This sometimes occurred with no change in the measured fracture stiffness.
Further investigation is needed, focusing on the visco-elastic properties of callus, to develop a more reliable method of determining clinical union.
Correspondence should be addressed to: LĂ©ana Fourie, CEO SAOA, PO Box 12918, Brandhof 9324 South Africa.