RADIOGRAPHIC EVALUATION OF A NON-MODULAR ACETABULAR CUP



Abstract

Press-fit acetabular reconstructions have become the standard THA; however, controversies remain. The purpose of this study was to critically evaluate serial radiographs for initial cup stability, i.e. gaps and signs of periacetabular interface changes for a porous tantalum monoblock socket.

A multicenter study evaluating 574 primary THRs (542 patients) performed by 9 surgeons at 7 hospitals, all with a monoblock cup without screws. Analyses included clinical outcomes and detailed 2-year minimum radiographic evaluation by one independent observer (mean follow-up, 33 months).

Complications included 9 intra-operative acetabular fractures. Among the 123 cases excluded from radiographic evaluations: deceased (19), lost-to-follow-up (8), 7 early revisions (recurrent dislocations (6) and one trauma-related loosening), and sepsis (3). Patient demographics (414 hips): mean age 65 years (19–93); 58 percent females. Baseline radiographs revealed 113 zones in 85 hips (21 percent) with acetabular gaps; 36 in zone I, 72 in zone II, and 5 in zone III. Of these radiolucencies, 57 zones were 1 mm or less and 56 zones ranged from 2 to 5 mm. At last follow-up, 64 hips (75 percent) had complete gap fill-in, including 100 percent of gaps greater than 3 mm.

There were no socket migrations, no evidence of lysis, no revisions for loosening, and no complete periacetabular interface radiolucencies. The fill-in of preexisting OA cysts and gaps is attributed to adequate initial stability and osteointegration into the porous tantalum. These results suggest that a monoblock cup without screws is an attractive option in THA.

Correspondence should be addressed to Richard Komistek, PhD, International Society for Technology in Arthroplasty, PO Box 6564, Auburn, CA 95604, USA. E-mail: ista@pacbell.net