CUT-OUT PHENOMENON AND OSTEOPOROSIS. RESULTS OF A PRELIMINARBIOMECHANICAL STUDY FROM EX VIVO HUMAN FEMORAL HEADS



Abstract

The failure rate of peritrochanteric fracture implants ranges from 5 to 20 % The aim of the present preliminary study is to evaluate the relationship between Singh index (SI), and failure of internal fixation (cut-out), examining bone mechanical properties from ex-vivo human femoral heads.

The failure rate of peritrochanteric fracture implants ranges from 5 to 20 % The aim of the present preliminary study is to evaluate the relationship between Singh index (SI), and failure of internal fixation (cut-out), examining bone mechanical properties from ex-vivo human femoral heads.

From a methodological point of view we decided to use SI on the basis of previous literature,, our data seems not to be influenced by the use of this methodological approach. > Three load speeds were set: 3, 10, 500 mm/min resulting values of BW/s were compared with Davy and Bergmann ones (obtained with instrumented prosthesis), and test validity was confirmed. In conclusion our data demonstrated that: cut out phenomenon occurs for lower mechanical load in femoral heads of patients with lower SI.