Abstract
The Self Aligning (SAL I) total knee prosthesis (Sulzer, Winterthur) is a first generation cruciate retaining mobile bearing implant allowing unrestricted rotation and up to 5 mm. of translation. A consecutive series of two hundred and three SAL I components in one hundred and sixty- seven patients were implanted between 1990 and 1994 and were reviewed clinically and radiographically. Quality of life outcomes (SF-12 and WOMAC) were also reviewed pre-operatively and regular intervals. This first generation mobile bearing knee implant in this series, continues to function well at a minimum ten years follow-up, with a mechanical failure rate (8%) comparable to fixed bearing designs.
A consecutive series of SAL I total knee arthroplasties were performed between 1990 and 1994 and were reviewed clinically, radiographically and with standardized quality of life outcome tools, including the SF-12 and the Western Ontario MacMaster Osteoarthritis Index (WOMAC).
Patient demographics, radiographs, and multiple validated outcome measures (WOMAC, SF12, Knee Society Clinical Rating System) were evaluated pre-operatively, at three, twelve, and twenty-four months and every two years thereafter.
The series included two hundred and three SAL I implants in one hundred and sixy-sevenpatients. There have been sixty-nine patient deaths (eighty-five procedures), nine patients (eleven procedures) now reside in nursing homes and are unable to return due to medical complications, two patients (three procedures) could not be located. Nine-teen patients/procedures have had revision surgery. Therefore, seventy-six patients (ninety-three procedures) were evaluated with over ten year follow-up (average 10.89 years, range 10.03–13.72). Indications for revision included, sepsis (three), aseptic loosening (eight), instability (two) and poly-wear (six). The KSCRS was assessed for all remaining patients at greater than ten years including pain (45.83), function(65.60), range of motion extension(0.13) and flexion(113.82) as well as the results of the WOMAC (pain-4.36, joint stiffness-2.29, function-18.66, total- 15.92) and SF-12 (Mental-55.37 Physical- 34.99).
The SAL I mobile bearing implant continues to perform comparably to fixed bearing cruciate retaining designs.
This first generation mobile bearing knee implant continues to function well at a minimum ten years follow-up, with a mechanical failure rate (8%) comparable to fixed bearing designs.
Correspondence should be addressed to Cynthia Vezina, Communications Manager, COA, 4150-360 Ste. Catherine St. West, Westmount, QC H3Z 2Y5, Canada