Abstract
This randomized clinical trial compares fixed- and mobile-bearing total knee prostheses in terms of the patients’ clinical outcome parameters (Knee Society Clinical Rating, WOMAC, SF-12), range of motion and performance during gait analysis for level-ground walking. Our results show no significant differences in the clinical outcomes and gait performance of the fixed- and mobile-bearing total knee arthroplasties.
The purpose of this study was to compare the clinical outcomes and gait parameters of patients with a fixed-bearing or mobile-bearing total knee arthroplasty (TKA).
Fifty-five patients were entered into a prospective, randomized clinical trial comparing fixed- versus mobile-bearing TKAs (Genesis II, Smith & Nephew, Memphis, TN). From this patient population, fifteen fixed-bearing and fifteen mobile-bearing TKA patients were matched based on age, sex and BMI to undergo gait analysis. Patients performed trials of level-ground walking at a self-selected velocity while three-dimensional kinetic and kinematic data were collected.
The fixed-bearing and mobile-bearing TKA patient groups were comparable regarding Knee Society Clinical Rating (181 ± 22 versus 171 ± 28), WOMAC scores (7 ± 5 versus 9 ± 12), SF-12 and range of motion (121° ± 11° versus 125° ± 6°).
Patients with fixed- and mobile-bearing TKAs performed similarly in the gait analysis in terms of their velocity, percent weight acceptance in the operated versus the non-operated limb, peak flexion in stance and swing phases, the support moments and extension moments at the ankle, knee and hip. Decreased peak extension in the mid-stance and swing phases was observed in the operative limb versus the non-operative limb for both fixed- and mobile-bearing TKAs (P=0.02 and 0.04). Decreased peak extension was also observed during mid-stance and swing phases in the mobile-bearing TKAs versus the fixed-bearing TKAs (P=0.064 and 0.052).
Fixed-bearing and mobile-bearing TKAs perform similarly in terms of their clinical outcome measures and the kinetics and kinematics of level-ground walking.
Funding for this project obtained from Smith & Nephew, Memphis, TN.
Correspondence should be addressed to Cynthia Vezina, Communications Manager, COA, 4150-360 Ste. Catherine St. West, Westmount, QC H3Z 2Y5, Canada