MINIMALLY-INVASIVE COMPUTER-ASSISTED HIP RESURFACING SURGERY USING THE ACROBOT® NAVIGATION SYSTEM



Abstract

This paper presents initial results of the Acrobot® Navigation System for Minimally Invasive (MI) Hip Resurfacing (HR) which addresses the problems of conventional HR. The system allows true MI HR – mini-mising the incision and tissue retraction required, and conservation of bone in contrast to other MI total hip procedures.

Pre-operative CT-based software allows the surgeon to plan the operation accurately. Use of CT gives the greatest accuracy, and is the only method which can give an accurate assessment of procedure outcome (planned versus achieved implant position). Intra-operatively, the bones are registered by touching points using a probe connected to a digitising arm. Next a series of tools is connected so that bone preparation and implant insertion is performed using on-screen guidance.

The accuracy of the registration probe is within 0.6mm, inside the acceptable margin for optical tracker systems. We have validated this acceptability using registration simulations leading to a protocol which restricts registration errors to within 1.5mm and three degree. These error margins are within those in the literature for acetabular component placement using optical tracker based systems (five degree inclination, six degree anteversion). No comparable data could be found regarding the accuracy of femoral component placement during computer-assisted HR.

The system is currently undergoing clinical tests at one alpha site, with three further beta sites planned for early 2006. The methods described by Henckel et al (CAOS International Proceedings 1994, pp. 281–282) are being used to evaluate the performance of the system, comparing pre-operative to post-operative CTs to obtain a true, accurate measure of performance.

Address for Correspondence: Mr K Deep, General Secretary CAOS UK, 82 Windmill Road, Gillingham, Kent ME7 5NX UK. E Mail: caosuk@gmail.com