Abstract
Introduction: Trabecular metal associated with monob-lock elliptical design represents a valid surgical solution for orthopaedic acetabular reconstructive procedures and second surgery.
Materials and methods: From 1999 to 2004, 61 patients between 45 and 81 years with osteoathritis underwent total hip primary arthroplasties with porous tantalum elliptical cup. We performed clinical evaluation through Harris Hip Score test. Bone-implant interface was studied through radiography with reference to the three Charnley’s areas. Follow-up were performed preoperatively at six months and yearly thereafter.
Results: Clinical results showed high improvement of Harris Hip Score (average preoperatively score was 46, postoperatively 90). Radiographic evaluation revealed a bone apposition to the porous tantalum without radiolucent lines around the acetabular interface. No local (osteolysis) or general (DVT) complication was seen.
Conclusions: Tantalum monoblock elliptical acetabular cup with high volumetric porosity, flexibility and high biocompatibility associated with particular microstructure permits direct apposition of bone, more extensive osseointegration with the maximum bone contact. The trabecular metal cup increases the initial stability helping in the prevention of osteolysis and loosening in a five years follow up.
Correspondence should be addressed to Ariella Neustadt at Studio EGA, Professional Congress Organisers, Viale Tiziano, 19 - 00196, Rome - Italy.