THE POSITIONAL MRI CHANGES IN THE LUMBAR SPINE FOLLOWING INSERTION OF A NOVEL INTER-SPINOUS PROCESS DISTRACTION DEVICE



Abstract

Introduction Symptoms of neurogenic intermittent claudication in spinal stenosis are explained by the narrowing of the spinal canal in the extended (upright) position and widening in the sitting (flexed) position. The XStop inter-spinous process distraction device is a new product that is designed to hold the affected segments in a flexed posture. This prospective study looks at the changes in the lumbar spine in a variety of postures from pre- to post insertion.

Methods Using a positional magnetic resonance imaging (pMRI) scanner, patients were scanned before and six months after the insertion of the device. Images were taken in sitting flexed and extended, and standing positions. The change in the total range of movement of the lumbar spine and in the individual operated segments was measured along with changes in the surface areas of the exit foramen, the dural sac, and the disc height.

Results 12 patients with 17 levels distracted have been scanned and measured. The cross sectional area of the dural sac at the level of the stenosis has increased from a mean of 77.8 mm2 to 93.4 mm2 in the standing position (p=0.006) and from 84.56mm2 to 107.35mm2 on extension (p=0.008). There were no statistically significant changes in the range of movement of the whole lumbar spine, or at levels adjacent to the device.

Discussion This study demonstrates that the X Stop device increases the cross sectional surface area of the spinal canal at the stenosed level, without causing extensive changes in the posture of the lumbar spine.

Please send correspondance to BASS at the Royal College of Surgeons, 35 - 43 Lincoln’s Inn Fields, London WC2A 3PE, UK.