Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

POSTERIOR SLOPE OF THE TIBIAL IMPLANT AND THE LONG TERM (16 YEARS) OUTCOME OF UNICOMPARTMENTAL KNEE ARTHROPLASTY

7th Congress of the European Federation of National Associations of Orthopaedics and Traumatology, Lisbon - 4-7 June, 2005



Abstract

The purpose was to assess the effect of the posterior slope on the long-term outcome of unicompartmental arthroplasty in knees with intact and deficient anterior cruciate ligaments.

We retrospectively reviewed ninety-nine unicompartmental arthroplasties after a mean duration of follow-up of sixteen years (12 to 20 years). At the time of the arthroplasty, the anterior cruciate ligament was considered to be normal in fifty knees, damaged in thirty-one, and absent in eighteen. At the most recent follow-up, we measured the posterior tibial slope and the anterior tibial translation on standing lateral radiographs.

In the group of seventy-seven knees that had not been revised by the time of the most recent follow-up, there was a significant linear relationship between anterior tibial translation (mean, 3.7 mm) and posterior tibial slope (mean, 4.3) (p = 0.01). The mean posterior slope of the tibial implant was significantly less in the group of seventy-seven knees without loosening of the implant than it was in the group of seventeen knees with loosening of the implant (p = 0.03). Five ruptures of the anterior cruciate ligament occurred in knees in which the ligament had been considered to be normal at the time of implantation; the posterior tibial slope in these five knees was greater than 13 degrees. Clinical evaluation revealed normal or nearly normal anteroposterior stability at the time of the most recent follow-up in all sixty-six unrevised knees in which the anterior cruciate ligament had been present at the time of implantation. Of the eighteen knees in which the anterior cruciate ligament had been absent at the time of the arthroplasty, eleven still had the implant in situ at the time of the most recent follow-up; the mean posterior tibial slope in these 11 knees was less than 5 degrees. Seven knees in which the anterior cruciate ligament had been absent at the time of the arthroplasty were revised. In these 7 knees, the tibial prosthesis was implanted with a posterior slope greater than 8 degrees.

These findings suggest that more than 7 degrees of posterior slope of the tibial implant should be avoided, particularly if the anterior cruciate ligament is absent at the time of implantation. An intact anterior cruciate ligament, even when partly degenerated, was associated with the maintenance of normal anteroposterior stability of the knee for an average of sixteen years following unicompartmental knee arthroplasty.

Theses abstracts were prepared by Professor Roger Lemaire. Correspondence should be addressed to EFORT Central Office, Freihofstrasse 22, CH-8700 Küsnacht, Switzerland.