Abstract
Introduction: The correct position of the knee arthroplasty components is associated with a better result of the prosthesis.
In the tibial component, both intramedullar and extramedullar instrumentations have been used for its fiability, but in the femoral component intramedullar guides are more precise than extramedullar ones.
The use of the intramedullar guide for the femoral component is not always possible, because a significant deformity of the femoral shaft or when a intramedullar device has been implanted in the femur.
We have studied the alineation of the components of computer assisted total knee arthroplasties in a group of patients with femoral deformities or implants.
Material and methods: We have used the surgical navigator Stryker-Howmedica for the implantation of a knee arthroplasty in a group of 10 patients in which a endomedullar femoral guide can not be used for femoral shaft severe deformities (6 cases): Paget disease (1 case), previous femoral osteomyelitis (2 cases) or previous femoral fractures (3 cases), or a shaft device was in the femoral shaft (4 cases): long hip femoral stem (3 cases) or a femoral nail (1 case) .
We have studied the alineation of femoral and tibial components with a whole-leg X-ray and Computer Tomography.
Results: All the femoral and tibial components have been implanted in a good position (90 +/– 2 degrees in the A-P plane and a femorotibial axe 180 +/– 3 degrees. The alineation in the sagital and axial planes have been inside the desired values in all cases also.
Discussion: It is generally accepted than intramedullary guides for the femoral component is the gold standard in arthroplasty of the knee.
In the last years, the development of computer assisted systems has allowed to obtain femoral and tibial cuts referred to the mechanical axes of the bone, without using mechanical guides for the alineation.
In some studies these navigation systems are better than mechanical instruments in terms of alineation of the components in cases without great deformities.
In this study, with some cases with severe femoral shaft deformities or with some intramedullary devices that does not allow the use of intramedullary femoral guides, we think that the indication to use a surgical navigator should be nearly absolute.
Theses abstracts were prepared by Professor Roger Lemaire. Correspondence should be addressed to EFORT Central Office, Freihofstrasse 22, CH-8700 Küsnacht, Switzerland.