Abstract
Introduction: Minimally invasive surgery (MIS) is associated with reduced postoperative morbidity and faster recovery of function. The PerCutaneous Compression Plate (PCCP) device was recently developed by Got-fried as a MIS technique for the fixation of osteoporotic hip fractures. PCCP provides rotational stability by means of two hip screws, and lateral cortical support by a proximal extension of the plate and by the relatively small diameter (9.3 mm) of the hip screws. The purpose of this prospective study is to compare the outcome of PCCP to the “gold standard” Compression Hip Screw (CHS) device.
Methods: 104 Patients with intertrochanteric fractures were randomized to be treated by PCCP (50 patients) or CHS (53 patients). One patient was switched from PCCP to CHS during surgery. Inclusion criteria were age above 60, close fracture reduction, no pathological fracture, and no surgical procedure in the same leg in the last year.
Results: The groups were comparable in patient age, gender, ASA, length of surgery and hospital stay. Operative blood loss was 177.8 ml in the PCCP group and 371.3 ml in the CHS group (p< 0.0001). At the 6th week clinic visit, patients in the PCCP group were able to bear more weight on the injured leg than patients in the CHS group (p< 0.03). Mortality during the first year follow-up period was 10% in the PCCP group and 24.5% in the CHS group (p~0.05). Analysis of X-ray radiographs revealed collapse in 4% of the patients in PCCP group and 19% in CHS group (p< 0.01).
Conclusions: Our results suggest that PCCP provides some of the advantages of MIS: reduced blood loss, as well as improves the stability of fracture fixation, demonstrated by improved early weight bearing and less fracture collapse. We found a trend for decreased first year mortality rate.
The abstracts were prepared by Ms Orah Naor. Correspondence should be addressed to Israel Orthopaedic Association at PO Box 7845, Haifa 31074, Israel.