Abstract
Introduction: Atypical femoral geometry and bone stock loss may make fitting a press-fit, traditional one-piece stem difficult, if not impossible. Modularity that permits independent fitting of the diaphysis, metaphysis, length, offset, and version may greatly facilitate reconstruction and ultimate clinical performance.
Method: We prospectively evaluated the early clinical results of Exactech’s AcuMatch®□ M-series prosthesis, a novel three-piece modular femoral stem. One hundred and twenty-seven were implanted in 114 patients. Ninety-one of these were primaries and 36 were revisions. The patients were followed prospectively and evaluated pre-operatively and at ≥2-year post-operatively.
Results: Harris hip score for primaries was 37 pre-operatively and 87 at last follow-up. Harris hip score for revision group was 35 pre-operatively and 84 at last follow-up. Four dislocations occurred and one prosthesis fractured in a 330-pound patient with deficient abductors. One prosthesis has been removed, due to sepsis in the primary group.
Conclusion: Through atraumatic milling of the bone and insertion of the implants, a minimal amount of intra-operative complications resulted. Ultimate restoration of hip biomechanics and soft tissue tension resulted in good clinical outcome with a very low dislocation rate.