DISTAL FEMORAL ENDOPROSTHETIC REPLACEMENT FOR NON-TUMOROUS CONDITIONS



Abstract

Introduction and Aims: Orthopaedic oncologists are often consulted regarding problems involving salvage of the distal femur due to bone loss, non-unions, infections. In young patients, extensive bony reconstruction is often necessary; in elderly, low demand patients, replacement with an endoprosthetic device results in decreased surgical time and more rapid mobilisation.

Method: Since 1991, 27 patients underwent reconstruction with a custom modular distal femoral replacement and rotating hinged knee joint (DFR). Twenty-two (81%) were revised to a DFR from an existing knee arthroplasty. Diagnoses included fracture, non-union, osteomyelitis, osteolysis or deformity. Average age was 66 (25–85); 83% were female. Most patients had undergone multiple prior surgeries. Patients with a history of infection had undergone aggressive resection and insertion of spacers with prolonged antibiotic administration, however they had no infection at the time of DFR reconstruction. All endoprostheses were cemented. Patients were allowed immediate weight-bearing and rehabilitation similar to patients undergoing TKA.

Results: One elderly patient died in the immediate peri-operative period of respiratory failure and one was lost to follow-up after placement in a nursing home. Average follow-up on 25 evaluable patients was 47 months (7–122). Reoperations were for recurrent infection (six) and tibial component loosening (three). Five of the six with infection were treated with synovectomy, antibiotic beads and suppressive oral antibiotics, and all five devices are still in place at an average of 54 months (range, 25–100). One severely diabetic patient had had multiple episodes of sepsis unrelated to the prosthesis which eventually seeded the distal femur and required a hip disarticulation. MSTS functional scores at last follow-up averaged 49% (13–80%) and HSS knee scores averaged 71% (37–90%).

Conclusion: DFR is a useful salvage procedure in low demand patients. Initially, six patients were scheduled for transfemoral amputation and three were confined to wheelchairs. Patients other than the hip disarticulation were at minimum household ambulators at last follow-up. In spite of problems with infection, most patients improved in overall function.

These abstracts were prepared by Editorial Secretary, George Sikorski. Correspondence should be addressed to Australian Orthopaedic Association, Ground Floor, The William Bland Centre, 229 Macquarie Street, Sydney, NSW 2000, Australia.

At least one of the authors is receiving or has received material benefits or support from a commercial source.