Osteolysis and subsequent mechanical loosening often occurs in hip arthroplasties using polyethylene-on-ceramic (POC) bearings. This has prompted an ongoing search for alternative bearing surfaces. Ceramic-on-ceramic (COC) and metal-on-metal (MOM) prostheses are widely used, with good clinical results. Using hip simulator studies, we compared ceramic-on-metal (COM) and MOM prostheses.
We found COM pairings had 100-fold lower wear rates than MOM. The wear particles from both articulations were oval to round in shape and in the nanometer size range, with the COM producing smaller particles than the MOM. In both pairings, particle size decreased as the bearings bedded in. The volumetric particle loads were far smaller with COM bearing-surfaces than in currently-used MOM prostheses.
These findings have encouraged us to investigate the use of these novel bearing surfaces. Ethical approval has been obtained, and a prospective randomised clinical trial comparing POC, MOM, COC and COM bearing surfaces has started.