Abstract
Study Design: Prospective observational study.
Objective: To establish the sensitivity, specificity and cost-effectiveness of motor evoked potential (MEP) monitoring of lower lumbar nerve roots during instrumented spinal fusion.
Subjects: 161 patients undergoing elective lumbar spinal fusion monitored with the Neurosign 800 machine.
Outcome Measures: MEP evidence of pedicle breaches and nerve root over-distraction. Symptoms and signs of new neurological deficits postoperatively. EMG confirmation of neurological deficits in symptomatic post-operative patients.
Results: True positive results consisted of pedicle breaches detected in 15 patients (9.3%). Nerve root irritation on distraction was found in 9 patients (5.6%). These results allowed modification of the surgical technique to prevent subsequent neural injury. True negative results on active pedicle probing occurred in 134 patients (83.2%) and in 146 patients (90.7%) on passive monitoring. False positive results were detected in 7 patients (4.3%). Four patients had electrical connection problems and in three patients pedicle probing was positive but direct screw testing was negative. True negative results consisted of a failure of monitoring to detect clinically significant neurological events in five patients (3.1%). In four the symptoms and signs were transient, resolving within six weeks of surgery. In one, revision decompression of the L5 nerve roots was required.
Conclusions: MEP monitoring in our hands has a specificity of 95.4% and a sensitivity of 75%. The cost per case is around £75.
These abstracts were prepared by Mr. Brian J C Freeman FRCS (Tr & Orth). Correspondence should be addressed to him at The Centre for Spinal Studies and Surgery, University Hospital, Queens Medical Centre, Nottingham NG7 2UH.