TOTAL HIP ARTHROPLASTY POLYETHYLENE WEAR AT TEN YEARS: INFLUENCE OF FEMORAL OFFSET



Abstract

Purpose: Polyethylene wear in total hip arthroplasty remains the most limiting factor for implant survival. Several predictive factors are well identified, but the position of the articulating pieces remains to be studied in detail. We searched for a correlation between polyethylene wear and the position of the femoral and acetabular pieces, particularly the femoral offset.

Material and methods: Sixty-six patients underwent total hip arthroplasty for osteoarthritis or osteonecrosis. The patients were reviewed at 10.8 years (four bilateral prostheses). The preoperative, immediate postoperative (1 month) and last follow-up (10 years) AP pelvis views were digitalized. A dedicated software traced the different axes for measurement. Wear at ten years, femoral offset, cup eccentration or medialisation, ascent or descent, and cup inclination were measured.

Results: Mean polyethylene wear was 1.23 mm at ten years with linear curve of 0.11 mm/yr. Preoperative femoral offset was restored in 71.4% of the cases. Univariate regression analysis revealed that only femoral offset was correlated with less wear at ten years. Polyethylene wear at ten years fell from 1.26 mm for preoperative offset restitution less than 98% to 1.13 mm for restitution greater than 102%.

Discussion: Image processing allowed greater accuracy in the measurement of polyethylene wear. The rate of wear reported in the literature ranges from 0.1 to .015 mm/yr. Restitution of femoral offset guarantees less wear due to the reduction in the resultant force applied on the articulation as well as stress on the implants. Furthermore hip stability is improved. Several factors are involved in production of wear debris and correct restitution of the centre of rotation is only one of the elements which reduce wear.

Conclusion: Wear was not excessive in this series. Among the position parameters, only femoral offset had an influence, having a beneficial effect on polyethylene wear. This emphasises the importance of having a wide variety of implants available in order to respond to the different anatomic presentations of the femur.