Abstract
Introduction and purpose: This is a prospective analysis of the clinical and radiographical results of two different types of PE, both associated to the same kind of prosthesis (Zimmer-Centerpulse).
Materials and methods: A total of 101 Allofit shells were analyzed; 56 had Sulene-PE (nitrogen sterilized) and 45 had Durasul highly cross-linked PE, associated to an Alloclassic stem (femoral head: 28 mm). They were all implanted between 1999 and 2002. Mean follow-up was 29.4 months for Sulene-PE and 25.3 for Durasul-PE. PE wear was assessed by means of image digitalization with special software (AutoCAD 14R) after six weeks (initial settling or position zero), after 6 and 12 months and on a yearly basis.
Results: Three cases dislocated and were thus excluded from the study. There were no infections. All cases assessed had good clinical and radiological results. There were no instances of implant loosening. There were no radiolucent lines, osteolysis, cortical hyperthrophy or proximal osteopenia. On the initial radiograph (position zero), the distance between the axes of the femoral head and those of the shell was 0.30+0.094 mm for the Sulene-PE group and 0.20+0.074 for the Durasul-PE one (p=0.029). Mean wear, taking position zero as a reference point, was 0.1035+0.0686 and 0.0819+0.078 (p=0.108) respectively.
Conclusions: In spite of the higher error rates found in the wear measurements of the Allofit cup, a greater position zero was found in Sulene-PE than in Durasul-PE. Although mean wear was higher in Sulene-PE than in Duarsul-PE, the differences found were not significant 3 years postop.
The abstracts were prepared by Dr. E. Carlos Rodríguez-Merchán, Editor-in-Chief of the Spanish Journal of Orthopaedics and Traumatology (Revista de Ortopedia y Traumatología). Correspondence should be sent to him at Sociedad Española de Cirugía Ortopédica y Traumatología (SECOT), Calle Fernández de los Ríos, 108, 28015-Madrid, Spain