Abstract
To determine whether increased sagittal laxity has an effect on functional outcome following posterior cruciate retaining total knee replacement using two differing tibial insert designs.
Ninety-seven patients were reviewed clinically, radiologically and underwent KT1000 testing of their TKR at a minimum follow up of 5 years (mean 6.5 yrs). The femoral component design was the same in all patients (Duracon/PCA). Fifty two patients had a relatively flat tibial insert design (group 1), while 45 patients had an AP lipped insert (group 2) following a change in design in 1995.
The 2 groups were comparable for age, sex, Charnley category, BMI, tibial slope and follow up. There was no significant difference in laxity measurements, IKS or WOMAC scores between the groups. There was no significant correlation between laxity and outcome score or flexion range.
Increased sagittal laxity in a knee replacement does not have a strong influence on functional outcome. The differing tibial insert designs had no significant effect on either laxity or function.
The abstracts were prepared by Jean-Claude Theis. Correspondence should be addressed to him at Department of Orthopaedic Surgery, Dunedin Hospital, Private Bag 1921, Dunedin, New Zealand.