CHARACTERIZATION OF TISSUES FROM METAL-METAL THRS



Abstract

A recent study of tissues from 14 modern metal-on-metal (MM) total hips reported an intense diffuse and perivascular (p.v.) lymphocytic infiltrate, suggestive of hypersensitivity (Willert et al. Osteologie 2000; 9:2–16). This study evaluated the histopathology of tissues from modern MMs using cases obtained at revision or autopsy.

Materials and methods: 35 MM THRs or surface replacements (SRs) that failed due to dislocation, aseptic loosening, and pain or obtained at autopsy (n = 4) were used. H& E stained sections were rated semiquantitatively. Selected cases were studied by immunohistochemistry for macrophage (CD68) and lymphocyte markers (CD3, 4, 20). Wear was measured with a coordinate measuring machine.

Results: Generally, the THRs without metallosis showed minimal visible wear particles, consistent with their low measured wear (av. total wear depth was 8.25 ± 6.7 um at av. 30 mos). Although SRs had an av. linear wear depth of 46 ± 48 microns at av. 23 mos, the metal rating was also low (av. 0.8), except in 1 case with HA 3rd body induced high wear and subsequent osteolysis. Lymphocytic aggregates were not a common feature but B type cells were extensive in 1 case (THR revised for pain after 36 months) moderate in 1 autopsy SR (with CoCr metallosis due to run-in wear of an out of round component) and minimal in 4 of the SRs.

Discussion and conclusions: Extensive diffuse or p.v. lymphocytes were not a consistent finding in these 35 cases. These features were not seen in well-functioning autopsy retrieved cases with low wear rates, nor in the SR with osteolysis and the highest amount of component wear. Until the long-term local and systemic effects of metal wear products, including hypersensitivity are better understood, continued histopathological assessment of periprosthetic tissues from MM total hips is recommended.

The abstracts were prepared by Nico Verdonschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, University Medical Centre, PO Box 9101, 6500 HB Nijmegen, The Netherlands.