Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

CEMENTLESS THA USING A CUSTOM-MADE BLASTED-SURFACE FEMORAL COMPONENT IN PATIENTS WITH SECONDARY OSTEOARTHRITIS OF THE HIP. FOUR- TO SEVEN-YEAR FOLLOW-UP STUDY.



Abstract

Introduction: The purpose of this study is to evaluate the clinical and radiogra phic outcome of patients with secondary osteoarthritis of the hip, who underwent custom-made cementless THA.

Methods: Between January 1994 and June 1997, 98 prim ary cementless custom-made THA’s were performed in 76 patients with secondary osteoarthritis and reviewed at mean 6 years follow-up (range: 4 – 7.5 years). Nine patients had a previous femoral osteotomy. There were 69 females and seven males. The mean age at operation was 54 years (40 – 73 years). Custom-made, 125 mm-long, titanium femoral components with blasted surface were fabricated based on the computerized tomography.

Results: The mean Harris Hip Score improved from 43 to 95 points. 97 patients (99%) had more than 80 points for total score, and 2 patients (2%) had thigh pain at the latest follow-up. Radiographically, 86 hips (87%) showed obviously extensive bone ongrowth onto the middle part of the stem while 8 hips (9%) showed stable fibrous fixation. Four hips (4%) were unstable: more than 2 mm subsidence in 2 hips and more than 2 degrees varus migration in 2 hips. Of these 4 hips, 2 had intraoperative cracks of the poximal femur and 2 had varus positioning of the stem.

Discussion and conclusion: Cementless THA has been advocated to enhance fit and fill of variable hip geometry. However, some clinical studies failed to show that custom implants significantly improve clinical success or implant longevity because their surface finish was not optimal. Custom-made titanium femoral components with blasted surface showed good clinical results. The reasons for radiolographical failures were because of intraoperative technical errors and these might be derived from 125 mm-long stem with maximum canal fill. A shorter stem may be an option because it may improve the feasibility of stem insertion and the distal fill.

The abstracts were prepared by Nico Verdonschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, University Medical Centre, PO Box 9101, 6500 HB Nijmegen, The Netherlands.