header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

ANTHROPOMETRY OF THE PROXIMAL TIBIA TO DESIGN TOTAL KNEE PROSTHESIS FOR THE JAPANESE POPULATION



Abstract

As for the number of patients who requires total knee arthroplasty (TKA), Asian-Pacific countries will be the most important market. However, due to the paucity of anthropometric data on the proximal tibia in this population, many prostheses designed for Caucasian knees have been introduced without specific modification.The aim of the current study was to analyze the geometry of the proximal tibia to design the optimum component for the Japanese population.

Anthropometric data on the proximal tibia of 100 knees in 80 patients undergoing TKA was obtained. Briefly, anterior-posterior (AP) and medial-lateral (ML) lengths were measured with a combination of two different methods, namely on the computed tomography (CT) images obtained preoperatively and intraoperative direct measurement on tibial resection surface. Reproducible measurement was possible only when the intraoperative measurement was combined with the corresponding CT images while the direction of measurements being aligned to the epicondylar axis of the femur.

It was shown that smaller components with an ML of around 60 mm were rarely required. Tibial component size variation should focus on an ML length of 65 to 75 mm because 76 of 100 knees (76 %) fell into this size range. When the subjects were confined to women, 70 of 77 knees (91%) were included in this size range. The intraoperative AP to ML ratio had a negative correlation with the ML length (r = −0.412, P < 0.0001) indicating that bigger knees were shallower in the AP direction. The size variation of currently popular pros-theses were not in accordance with the geometry of the tibial resection surface shown in this study.

The results of this study applied to a cross-section of the Japanese population can be used by manufacturers to create a prosthesis suitable for most of the Asian-Pacific population.

The abstracts were prepared by Nico Verdonschot. Correspondence should be addressed to him at Orthopaedic Research Laboratory, University Medical Centre, PO Box 9101, 6500 HB Nijmegen, The Netherlands.