Advertisement for orthosearch.org.uk
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

O3244 COMPUTER-ASSISTED CUP PLACEMENT IN TOTAL HIP ARTHROPLASTY:WHICH PRECISION AND ACCURACY?



Abstract

Aims: To determine the precision of conventional versus computer assisted techniques for positioning the acetab-ular component in total hip arthroplasty (THA). Methods: Using a lateral approach, 150 cups were placed by 10 different surgeons in 10 identical plastic pelvis models. Only the immediate operating þeld was visible. Preoperative planning was performed with a computerized tomographic scan. Fifty cups were placed free hand, 50 others with the standard cup ancillary, and the remaining 50 cups using computer-assisted orthopaedic surgery. The accuracy of cup abduction and ante-version was assessed with an electromagnetic system. Results: Using conventional techniques, free hand placement revealed a mean precision of cup anteversion and abduction of 10¡ [range: 5.5–14] and 3.5¡ [2.5–5] respectively. With the cup positioner, these angles measured 8¡ [5–10.5] and 4¡ [3–5.5] respectively, and using the computer assisted method, the mean cup anteversion precision was 1.5¡ [1–2] and mean cup abduction measured 2.5¡ [2–3.5]. Conclusions: Computer assisted cup placement is a very accurate and reproducible technique during THA. It is clearly more precise than either of the two traditional methods of cup positioning, even for well-trained surgeons.

Theses abstracts were prepared by Professor Dr. Frantz Langlais. Correspondence should be addressed to him at EFORT Central Office, Freihofstrasse 22, CH-8700 Küsnacht, Switzerland.