SHOULD THE AO/ASIF DYNAMIC CONDYLAR SCREW BE USED IN THE TREATMENT OF PATHOLOGICAL SUBTROCHANTERIC FRACTURES OF THE FEMUR? A RETROSPECTIVE REVIEW OF 14 CASES AND A SUGGESTION FOR CAUTION WITH THIS DEVICE



Abstract

To determine whether the Dynamic Condylar Screw, DCS, is suitable in treating pathological subtrochanteric fractures of the femur and the incidence of failure with this device.

Nineteen sequential and unselected patients with twenty femurs with pathological subtrochanteric fractures, or impending fractures, were identified, which had been treated with DCS. Mean age was seventy. Follow up was until functional union (minimum follow up 18 months) or until death. Primary tumour was identified in seventeen of the nineteen cases.

Grade of surgeon was recorded as was the use of adjunctive measures at the time of surgery (PMMA). Subsequent failure, cause and need for revision was noted.

Operative morbidity was low and initial pain relief good in all cases. Two fractures went on to unite. Eight original implants survived until the patient died (mean survival time 24 days). Ten implants failed (50%). Failure resulted from fracture through the DCS plate in eight cases and cut out of the screw in the other two. Of these, five patients (6 femurs) went on to further operative procedures. The others were either deemed unfit (3 cases) or refused further surgery (1 case).

In the light of newer intramedullary techniques we feel the DCS should be used with caution in this type of fracture. DCS in this fracture is associated with a high degree of failure.

The abstracts were prepared by Mr Simon Donell. Correspondence should be addressed to him at the Department of Orthopaedics, Norfolk & Norwich Hospital, Level 4, Centre Block, Colney Lane, Norwich NR4 7UY, United Kingdom