header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

QUANTIFICATION OF IN VIVO LOADING IN THE ROTATOR CUFF



Abstract

In vivo loading data for the rotator cuff would be of value to scientists and clinicians interested in the shoulder in the testing of surgical repairs, design of rehabilitation programs and for finite element models.

A technique for insertion and retrieval of the Arthroscopically Insertable Force Probe (AIFP - Microstrain Inc. Burlington, Vermont, USA) from the subscapularis is described was initially established in a cadaveric model. Ethical approval was obtained for AIFP insertion into the subscapularis tendon of volunteers during diagnostic shoulder arthroscopy. An in situ calibration was carried out using a modified arthroscopic grasper ( Smith and Nephew, Huntingdon, UK). After motor effects of interscalene block had worn off dynamic data relating to subscapularis tendon loading was collected. The AIFPs were removed through a port site by traction on an O (3.5 metric) nylon suture without complication.

Maximum loading of the subscapularis tendon was measured during internal rotation from neutral with the arm fully adducted. Forces measured exceeded 200N.

This paper describes a novel technique for the insertion, calibration and retrieval of AIFPs from the rotator cuff. In vivo tendon loading data was obtained. The techniques described may be applied to other structures of interest to orthopaedic surgeons.

The abstracts were prepared by David Stanley. Correspondence should be addressed to him c/o British Orthopaedic Association, Royal College of Surgeons, 35–43 Lincoln’s Inn Fields, London WC2A 3PN.