header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

CONTROL OF JOINT CENTRE AND HORIZONTAL OFFSET USING A CUSTOM FEMORAL COMPONENT WITH A DURALOC SOCKET



Abstract

To maximise the long-term survivorship of any hip prosthesis it is important to recreate joint centre. Normal joint centre is determined by horizontal offset and vertical height of the acetabular and femoral components. In this study joint centre and horizontal offset were analysed in 200 consecutive patients operated on from October 1998 in whom the opposite hip was normal. Joint centre was defined relative to the acetabulum and femur both pre- and post-operatively. On the acetabular side a horizontal line was drawn across the pelvis immediately below each teardrop. A vertical line was drawn at right angles through the middle of each teardrop. Acetabular offset was defined as the horizontal distance from the vertical trans teardrop line to head centre. For femoral offset a screened x-ray was taken to show maximum offset. The anatomical axis was drawn and the offset was defined as the distance from the anatomical axis to head centre.

Our results show on the acetabular side there was an overall tendency to leave the joint centre medial and so decrease acetabular offset. However, we found that 90% of our sockets were placed within 6 mm of normal joint centre. We attribute this accuracy to the principle of visualising the transverse acetabular ligament intra-operatively and using this landmark to control depth of socket insertion. Conversely, on the femoral side there was a slight tendency to increase the offset. Nevertheless, 98% of the custom stems were within 10mm of normal joint centre. When we looked at total horizontal offset i.e. the combination of femoral and acetabular offset we found that joint centre had been restored to within 10mm in 93% of cases.

This study confirms the effectiveness of the custom femoral stem and Duraloc socket in restoring joint centre.

These abstracts were prepared by Mr Peter Kay. Correspondence should be addressed to him at The Hip Centre, Wrightington Hospital, Appley Bridge, Wigan, Lancashire WN6 9EP.