Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Femoral neck fracture fixation. Comparison of a sliding screw with lag screws



Download PDF

Abstract

The rigidity of a sliding compression screw and three cannulated lag screws in the treatment of subcapital fractures was compared in five pairs of female cadaver femora. There were no significant differences between the compressive strength, bone density, cortical thickness or Singh index of the bones in each pair. A subcapital fracture was standardised using a perpendicular saw cut across the femoral neck. A uniaxial 'load test system' with force and length measurement facilities was used to mimic cyclical stressing applied in vivo at a frequency of 0.5 Hz from 0 to 3 times body-weight. There was no significant difference between the fixation afforded by the sliding compression screw and three lag screws. Bone quality was the single most important factor in the stability of the bone implant unit.

For access options please click here