Cementless uncoated polyethylene acetabular components in total hip replacement. Review of five- to 10-year results



Download PDF

Abstract

We reviewed the results of 545 consecutive total hip replacements using a cementless non-coated high-density polyethylene acetabular component combined with a cemented Muller stem at five to 10 years. In all, 421 patients (445 hips) were available for review, 118 by questionnaire and 303 by examination and radiography. Of these, 86% had a good or excellent result. We found a high rate of radiological loosening of the cup after the sixth year, and a high rate of clinical loosening after the eighth year. Loosening was commoner in women, in younger patients and where a smaller size of acetabulum had been used. Calcar resorption was significantly related to loosening of the acetabulum. Loosening appeared to be mainly due to polyethylene debris produced by micro-movement of the acetabulum against the bone, which had resulted in a giant cell foreign body reaction and subsequent bone erosion. We have abandoned the use of this prosthesis and suggest that direct contact between bone and polyethylene should be prevented by a coating of metal or some other material.