The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 94 - 101
1 Jun 2021
Roy ME Whiteside LA Ly KK Gauvain MJ

Aims

The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components.

Methods

Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 116 - 122
1 Jun 2019
Whiteside LA Roy ME

Aims

The aims of this study were to assess the exposure and preservation of the abductor mechanism during primary total hip arthroplasty (THA) using the posterior approach, and to evaluate gluteus maximus transfer to restore abductor function of chronically avulsed gluteus medius and minimus.

Patients and Methods

A total of 519 patients (525 hips) underwent primary THA using the posterior approach, between 2009 and 2013. The patients were reviewed preoperatively and at two and five years postoperatively. Three patients had mild acute laceration of the gluteus medius caused by retraction. A total of 54 patients had mild chronic damage to the tendon (not caused by exposure), which was repaired with sutures through drill holes in the greater trochanter. A total of 41 patients had severe damage with major avulsion of the gluteus medius and minimus muscles, which was repaired with sutures through bone and a gluteus maximus flap transfer to the greater trochanter.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 23 - 26
1 Jan 2016
Whiteside LA

An extensive review of the spinal and arthroplasty literature was undertaken to evaluate the effectiveness of local antibiotic irrigation during surgery. The efficacy of antibiotic irrigation for the prevention of acute post-operative infection after total joint arthroplasty was evaluated retrospectively in 2293 arthroplasties (1990 patients) between January 2004 and December 2013. The mean follow-up was 73 months (20 to 139). One surgeon performed all the procedures with minimal post-operative infection.

The intra-operative protocol included an irrigation solution of normal saline with vancomycin 1000 mg/l and polymyxin 250 000 units/l at the rate of 2 l per hour. No patient required re-admission for primary infection or further antibiotic treatment. Two morbidly obese patients (two total hip arthroplasties) developed subcutaneous fat necrosis requiring debridement and one was revised because the deep capsular sutures were contaminated by the draining subcutaneous haematoma. One patient who had undergone total knee arthroplasty had unrecognised damage to the lateral superior geniculate artery and developed a haematoma that became infected secondarily four months after the surgery and underwent revision.

The use of antibiotic irrigation during arthroplasty surgery has been highly effective for the prevention of infection in the author’s practice. However, it should be understood that any routine prophylactic use of antibiotics may result in resistant organisms, and the wise stewardship of the use of antibiotics is an important part of surgical practice.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):23–6.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 31 - 36
1 Jan 2016
Whiteside LA Roy ME Nayfeh TA

Bactericidal levels of antibiotics are difficult to achieve in infected total joint arthroplasty when intravenous antibiotics or antibiotic-loaded cement spacers are used, but intra-articular (IA) delivery of antibiotics has been effective in several studies. This paper describes a protocol for IA delivery of antibiotics in infected knee arthroplasty, and summarises the results of a pharmacokinetic study and two clinical follow-up studies of especially difficult groups: methicillin-resistant Staphylococcus aureus and failed two-stage revision. In the pharmacokinetic study, the mean synovial vancomycin peak level was 9242 (3956 to 32 150; sd 7608 μg/mL) among the 11 patients studied. Serum trough level ranged from 4.2 to 25.2 μg/mL (mean, 12.3 μg/mL; average of 9.6% of the joint trough value), which exceeded minimal inhibitory concentration. The success rate exceeded 95% in the two clinical groups. IA delivery of antibiotics is shown to be safe and effective, and is now the first option for treatment of infected total joint arthroplasty in our institution.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):31–6.