Aims. We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. Patients and Methods. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy. Results. After adjustment for other factors, CoCr stem tapers were found
to have significantly greater volumetric material loss than the
equivalent Ti stem tapers. Conclusion. When taper
Aims. Asphericity of the femoral head-neck
This study was undertaken to elucidate the mechanism of biological repair at the tendon-bone
Traumatic unilateral facet dislocation of the lumbosacral
Aims. We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. . Patients and Methods. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper
We used Laser Doppler flowmetry to measure the effect on the blood flow to the femoral head/neck
Aims. Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper
Impingement by prominence at the femoral head-neck
We present a series of 35 patients (19 men and
16 women) with a mean age of 64 years (36.7 to 75.9), who underwent
total hip replacement using the ESKA dual-modular short stem with
metal on-polyethylene bearing surfaces. This implant has a modular
neck section in addition to the modular head. Of these patients,
three presented with increasing post-operative pain due to pseudotumour
formation that resulted from corrosion at the modular neck-stem
junction. These patients underwent further surgery and aseptic lymphocytic
vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the
neck-stem taper. Blood cobalt and chromium levels were measured
at a mean of nine months (3 to 28) following surgery. These were
compared with the levels in seven control patients (three men and
four women) with a mean age of 53.4 years (32.1 to 64.1), who had
an identical prosthesis and articulation but with a prosthesis that
had no modularity at neck-stem
Tuberculosis of the craniocervical
We undertook a retrospective study investigating
the accuracy and safety of percutaneous pedicle screws placed under
fluoroscopic guidance in the lumbosacral
Dislocations of the cervicothoracic
We present the case of an 83-year-old man who developed quadriparesis and respiratory embarrassment following osteomyelitis at the occipito-atlantoaxial
The early failure and revision of bimodular primary
total hip arthroplasty prostheses requires the identification of the
risk factors for material loss and wear at the taper
1. The causes of pressure on the neuro-vascular bundle of the upper limb are many and varied. No one cause such as clavicular pressure can explain all cases. 2. Costo-clavicular pressure is not possible when there is a normal first rib and a normal thoracic outlet, but it is certainly a factor when the costo-clavicular interval is narrowed by the presence of a large cervical rib or an abnormal first thoracic rib. Clavicular pressure can act only during retraction and abduction, not in depression of the shoulder. 3. Temporary alterations in the radial pulse on movements of the shoulder in normal individuals are due to causes distal to the clavicle and have no relation to costo-clavicular pressure. 4. While irritation of sympathetic nerve fibres may explain the majority of cases of thrombosis, there are others in which clotting occurs in an aneurismal dilatation produced by pressure between the clavicle and the abnormal costal element. It is likely that the thrombosis occurs in an aneurism which has been present for some length of time. The cause of the aneurismal dilatation may be vaso-motor paralysis of a segment of the artery, ending distally at a point where a fresh intact leash of nerves is relayed to the vessel. 5. The importance of the scalenus anterior syndrome has been over-emphasised. If operative treatment is limited in all cases to anterior scalenotomy the results will be disappointing. 6. If operation is advised it should be performed without rigid and preconceived ideas, through an adequate incision, and with exploration wide enough to allow thorough investigation of the cause of pressure.
Of 37 consecutive patients suffering from tuberculosis of the lumbosacral spine, 26 were reviewed after an average follow-up of 20 years. Those presenting at under 10 years of age usually had a discharging sinus or a pointing abscess and a visible kyphosis. In older patients, low back pain was the predominant presenting symptom. Neurological involvement was uncommon. Both operative and conservative treatment had resulted in fusion but all patients treated conservatively ended up with a kyphosis associated with trunk shortening. They had a higher incidence of back pain and more complications in pregnancy. Anterior debridement and fusion with a strut graft can reduce the incidence and size of kyphosis, but is technically demanding. Treatment remains difficult and requires individual consideration.
Aims. Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. Methods. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic
Aims. Whether a combined anteroposterior fusion or a posterior-only fusion is more effective in the management of patients with Scheuermann’s kyphosis remains controversial. The aim of this study was to compare the radiological and clinical outcomes of these surgical approaches, and to evaluate the postoperative complications with the hypothesis that proximal
Aims. Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. Methods. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the
Aims. The aim of this study was to investigate the incidence and characteristics of instrumentation failure (IF) after total en bloc spondylectomy (TES), and to analyze risk factors for IF. Methods. The medical records from 136 patients (65 male, 71 female) with a mean age of 52.7 years (14 to 80) who underwent TES were retrospectively reviewed. The mean follow-up period was 101 months (36 to 232). Analyzed factors included incidence of IF, age, sex, BMI, history of chemotherapy or radiotherapy, tumour histology (primary or metastasis; benign or malignant), surgical approach (posterior or combined), tumour location (thoracic or lumbar;