PERFORMANCE OF SYNOVIAL FLUID D-LACTATE TEST FOR ACCURATE DIAGNOSIS OF PERIPROSTHETIC JOINT INFECTION

European Bone and Joint Infection Society (EBJIS) meeting, Antwerp, Belgium, September 2019.



Abstract

Aim

To evaluate the analytical performance of synovial fluid D-lactate test for the diagnosis of PJI.

Method

Consecutive patients undergoing diagnostic joint aspiration of prosthetic joint were prospectively included. PJI was diagnosed according to the proposed European Bone and Joint Infection Society (EBJIS) definition criteria. Synovial fluid was collected for culture, D-lactate measurement (by spectrophotometry, λ = 570 nm) and leukocyte count and differential (by flow cytometry). The receiver operating characteristic (ROC) analysis was performed to assess the diagnostic performance of D-lactate and leukocyte count.

Results

Diagnostic joint aspiration was performed in 224 patients with prosthetic joints. PJI was diagnosed in 87 patients (39%). The optimal D-lactate cut-off value for diagnosing PJI was 1.2 mmol/l. The sensitivity of synovial fluid D-lactate was 97.7%, specificity 83.9%, whereas the sensitivity of synovial fluid leukocyte count was 87.5% with specificity 95.7%. Concentration of SF D-lactate was significantly higher in patients with PJI compared to aseptic loosening of prosthesis (median (range)) 2.33 (0.99–3.36) vs 0.77 (0.01–2.4), p<0.0001. We found positive correlation between D-lactate and erythrocytes in synovial fluid sample in the aseptic group (ρ = 0.339, p< 0.01).

Conclusions

The synovial fluid D-lactate showed a good diagnostic performance for the diagnosis of PJI, which was comparable to the synovial fluid leukocyte count. Currently available (UV)-based method for detection of D-lactate showed low specificity (84%) due to influence of hemoglobin with the similar absorbance wavelengths (λ = 540 nm). More specific high-performance methods such as electro-chemical sensing system or lateral flow immunochromatographic assays should be implemented.