Abstract
The MAGnetic Expansion Control (MAGEC) system is used increasingly in the management of early-onset scoliosis. Good results have been published, but there have been recent reports identifying implant failures that may be associated with significant metallosis surrounding the implants. This article aims to present the current knowledge regarding the performance of this implant, and the potential implications and strategies that may be employed to identify and limit any problems.
We urge surgeons to apply caution to patient and construct selection; engage in prospective patient registration using a spine registry; ensure close clinical monitoring until growth has ceased; and send all explanted MAGEC rods for independent analysis.
The MAGEC system may be a good instrumentation system for the treatment of early-onset scoliosis. However, it is innovative and like all new technology, especially when deployed in a paediatric population, robust systems to assess long-term outcome are required to ensure that patient safety is maintained.
Cite this article: Bone Joint J 2017;99-B:708–13.
The treatment of young patients with progressive early onset scoliosis necessitates the control of the deformity while permitting continued growth. This may be achieved using several methods including distraction-based techniques.1 Traditional growing rod systems involve repeated open operations to lengthen the rods. These are commonly associated with complications including deep infection and implant failure, and a psychological toll on the child and their parents.2-5
In 2009, a novel spinal implant, the MAGnetic Expansion Control (MAGEC) system (Nuvasive Specialised Orthopaedics, San Diego, Californa), received its European Conformity mark, allowing sales in the United Kingdom and Europe. This system is attractive in the management of early-onset scoliosis as it includes a magnetically driven linear actuator allowing non-invasive lengthening, potentially avoiding repeated operations and their associated complications. Given these possible benefits, the National Institute for Health and Care Excellence (NICE) in the United Kingdom recommended that the system should be considered in children with early-onset scoliosis as its use was supported by the evidence.6 It has subsequently been widely used in the United Kingdom. Approval by the Food and Drug Administration (FDA) has also been granted in the United States. The Bone & Joint Journal recently reported the experience with this implant from Cardiff, noting a high rate of failure, significant tissue metallosis surrounding the implants at the time of revision associated with metal debris, and fractures of a drive pin within the actuator.7 Such findings necessitate an evaluation of our current knowledge about this device and the mechanisms in place that are designed to detect failures.
Any system spanning unfused spinal levels will be liable to fracture and anchor failure, because of continuing movement of the spine. It is currently unclear, however, if the complex internal design of the MAGEC system is associated with significantly more complications as suggested by the work of Teoh et al.7 Cases of failure of the actuator mechanism through fracture of a drive pin have been reported from other centres.8-10 Metallosis has also been noted at the time of revision in two other reports.8,11 To date, there remains insufficient evidence in the literature to determine if such failures are common, as there are few long-term outcome studies following the use of the MAGEC system. Hosseini et al11 reported the outcome in 23 patients at a mean follow-up of two years. Complications necessitating re-operation occurred in ten patients. Six underwent revision of the rods themselves, mainly for breakage in two and problems with the anchorage in two; metallosis and failure of lengthening necessitated revision in the remaining patients.11 The Cardiff group12 have previously reported the outcomes of a slightly larger cohort of eight patients treated with this system, with a mean follow-up of four years. Complications necessitated re-operation in six patients; one for failure of lengthening and another for fracture of a drive pin. Fractures of a rod in two patients, failures of anchorage and proximal junctional failure prompted revision in the remaining patients.12 In a multicentre study involving 33 patients, ten revision procedures were required during the first 12 months post-operatively, including for ‘rod breakage’ and failure of lengthening in four and two patients respectively.13 La Rosa et al10 reported a fracture of a rod and one of a drive pin in ten patients, with a mean follow-up of 27 months. Two fractures of a drive pin were seen in nine patients at a mean follow-up of 29 months.8 In a multicentre French study of 30 cases, 24 complications occurred including six failures of the MAGEC rods to lengthen at a mean of 18 months follow-up. A total of 13 revisions had occurred in nine patients including three for ‘rod breakage’ and two for non-functioning rods.14 Other reports are more reassuring, with Thompson et al15 reporting proximal failure of anchorage in three of 19 patients, but no failure of a MAGEC rod itself at a mean follow-up of 22 months. Akbarnia et al16 reported on 12 patients with a mean follow-up of 2.5 years; four had undergone a re-operation but only one was for failure of a rod with loss of distraction. Dannawi et al17 encountered two ‘rod breakages’ in 34 patients followed for a mean of 15 months. A single centre Turkish study reported two fractures of a rod in 18 patients at a mean follow-up of 18 months.18
Given the contradictory and limited nature of the literature to date, further prospective studies reporting the long-term outcomes following the use of the MAGEC system in larger numbers are urgently needed.
Following reports of early failures of the MAGEC system, the British Scoliosis Society (BSS) audited its members in 2016. Data were received from 11 of 17 centres using the device and included 195 patients (369 rods) treated since its introduction onto the United Kingdom market. Unplanned revision surgery occurred in 43 patients (22%). Fracture of a rod occurred in 11(6%), 13 (7%) had a fracture of a drive pin and 14 (7%) had failure of lengthening without an obvious fracture of a drive pin. Metallosis was noted in ten patients (23%) undergoing unplanned revision surgery. Following the original audit, more recent data have been received from three of the 11 centres, noting four more cases of fracture of a drive pin in their original cohort giving an incidence of failure of a drive pin of 8%. Similar investigations are being undertaken by the European Paediatric Orthopaedic Society which plans to survey its members in 2017 to identify patients treated with the MAGEC system and to assess the clinical outcomes.19
The implications to patients and healthcare providers if the MAGEC system proves to perform poorly are significant. Currently, about 125 patients are treated with this system annually in the United Kingdom. It is difficult to estimate the number of patients treated worldwide. It is likely to be quite extensively used in the United States. Whilst the local effects of metallosis are striking in the images provided by Teoh et al,7 it is the systemic effects of metal debris that are potentially the most worrying. It should be noted that, like most traditional growing rods, the MAGEC system is made predominantly from titanium alloy (TiAl6V4), with an increase in the serum level of titanium associated with their use.20 Patients treated with the MAGEC system have also been shown to have significantly higher serum levels of vanadium than those treated with conventional systems.20 The long-term effects of raised levels of metal ions in children is unknown. Most studies have focused on the potential carcinogenesis related to cobalt/chrome based on arthroplasties in adults.21 Wear debris from arthroplasties containing titanium has been shown to induce aneuploidy in vitro and in vivo in a dose dependent manner.22
Any necessary revision surgery carries a burden of morbidity for the child but also significant financial implications. Despite a pair of MAGEC rods costing around £14 000, the initial cost analyses, available at the time of NICE approval, suggested a saving after three years compared with conventional growing rods.23,24 These cost analyses and national guidance may need to be revisited once the true survival of the implant becomes clear.
Over the last 20 years, patients have repeatedly suffered as a result of poorly performing orthopaedic implants, such as the 3M Capital hip (3M Health Care Ltd., Loughborough, United Kingdom).25 A novel implant may be used for many patients before adequate clinical data, upon which to assess its suitability for widespread use, becomes available. In response to the failures of previous implants, safer strategies for the introduction of new technology have been suggested, but none have been implemented.
The stepwise approach described by Malchau et al26,27 is conceptually attractive. The first step is to identify the incidence and severity of the problem. The incidence of the early-onset of scoliosis is low but the effects on the patient are severe. The second step proposes a solution, in this case MAGEC. The third step defines the extent of the ‘Universal Dilemma’, which is the gap between the pre-clinical results (e.g. animal or biomechanical testing) and actual clinical outcomes.26 This gap has yet to be defined for the MAGEC system. The fourth step is to identify the compromises driven by this dilemma. This might now be coming to light with the various reports of failure that have been published. The fifth and final step is to define the cost-effectiveness. NICE relied on data suggesting significant cost-effectiveness in the context of clinical effectiveness (‘clinical proficiency’)28 when it recommended the use of MAGEC for the treatment of early-onset scoliosis. That consideration may have to be reconsidered in the light of new data.
More recently, the IDEAL Framework (idea, development, exploration, assessment, long-term study) was developed as a rational mechanism for reducing the morbidity associated with poorly performing novel implants.29-31 Broadly, this system and the stepwise approach of Malchau,27 suggest initial evaluation with small prospective trials, followed by large, multicentre, ideally randomised, prospective studies, before the widespread adoption of an implant. Long-term monitoring using registries would form an integral part of that evaluation.28-31
In the absence of the application of these logical approaches in the United Kingdom, the optional ‘Beyond Compliance’ initiative potentially offers an improvement on the current practice of the introduction of a new implant, but at present it includes only hip and knee arthroplasties and its impact remains limited.32 Implants continue to be introduced through ‘substantial equivalence’ routes, avoiding the need for large-scale trials before widespread introduction. At the time of evaluation by NICE in the United Kingdom, only three peer reviewed case series of patients treated with the MAGEC system were available, and only two patients with follow-up of more than two years had been described.17,24,33-35 In the United States, the MAGEC system was cleared by the FDA via the 510(k) route. To do so a device must have the same intended use and be “at least as safe and effective, that is, substantially equivalent, to a legally marketed device” (i.e. predicate device). Technological characteristics employed must be the same as the predicate device or if different “does not raise new questions of safety and effectiveness”.36 In this case the MAGEC system was felt ‘substantially equivalent’ to the Harrington rod system, which involved static distraction instrumentation and had been introduced into clinical practice in the 1950s.37 The MAGEC actuator mechanism had been previously approved by the FDA when used within intramedullary nails,38 i.e. the Harrington Rods and intramedullary nails were considered to be predicate devices.39 Such judgments are surprising as it is established from other orthopaedic implants that even minor changes can lead to morbidity through unforeseen consequences.40
As a result of the current ways in which new implants are introduced into clinical practice, we rely heavily on registries for post-market surveillance.41 Registries have proved to be valuable in this role, recognising both poorly performing joint arthroplasties and bone cement.42 Most recently, the Australian and United Kingdom National Joint Registries identified the Articular Surface Replacement hip as an outlier, leading to its withdrawal.43,44 Using registries as a means of post-market surveillance is also favoured by the FDA.45 The British Spine Registry (BSR)46 was established in 2012 with goals that included improving patient safety and monitoring the outcomes of surgery. Recognising the lack of long-term clinical data at the time of evaluation of the MAGEC system, NICE suggested that appropriate data be captured using a registry.6 Furthermore, NHS England mandated the registration of all patients treated with the MAGEC system, within the United Kingdom, on the BSR.47
We have assessed the compliance with BSR registration of units treating patients with the system. The hospital and date of surgery for all patients treated with it in a NHS centre between 01 April 2014 and 31 March 2016 was provided by the manufacturer. These data were cross-referenced with the BSR. Data were analysed to assess the number of patients treated in each centre and compliance with registration in the BSR. Over the course of the study, 234 operations were performed in 19 NHS centres. National usage was consistent during this time with a mean of 29.3 cases per quarter. There was a wide variation in the number of patients treated with the system by different centres, with a median of seven patients/unit (interquartile range 4.0 to 16.5). Only 65 of the 234 patients (28%) were entered onto the BSR. The rate of entry varied considerably between centres (0% to 100%). Six (32%) made no entries. No unit that had treated more than four patients had 100% BSR entry. There was no significant change in the proportion of patients registered on the BSR per quarter during the study. As a result of the failure to comply with the NHS England mandate, current data for the MAGEC system in the United Kingdom is inadequate and cannot be used to identify the rates of re-operation after the use of the MAGEC system. Feedback has been provided to centres with poor compliance with the registry and they have been encouraged to place all patients on the BSR retrospectively and to record complications prospectively. Future work will identify whether this information will lead to a change in practice.
It has been suggested that a fully operational registry should capture > 80% of the national data. The rate of regi-stration of 28% of patients treated with the MAGEC system on the BSR is unacceptable.48,49 Such poor quality data is not only seen in the BSR. The Canadian Spine Outcomes and Research Network also reports significant variation in the rates of registration of patients between centres.50Follow--up is also poor in most other national registries, with only two of the 13 spine registries examined by McGirt et al49 reporting follow up of > 75% of patients, one year post-operatively. One such successful registry, SweSpine, was evaluated by Fritzell et al,51 who identified factors thought influential in its success. Considering these factors, it seems likely that greater commitment by surgeons, hospitals and regulators is required to improve compliance.
The importance of registries should be clear to all those responsible for the care of these patients, with the clinical necessity and mandate from NHS England for registration of MAGEC cases undeniable. There is an international applicability to this concept, not only for MAGEC, but also for other innovative products that might have long-term clinical consequences. Despite this, logistical support is a common obstacle. One centre in the United Kingdom reporting their experience of implementing the registry concluded “the support for BSR in NHS hospitals is woefully inadequate”.52Surgeons must work with their hospitals to facilitate both the resources and time to input data. The cost of these provisions was estimated by an institutional spine registry in the United States to be $160 per patient.49 As the cost of the implants alone is tens of thousands of pounds, euros or dollars, such a small sum is a pittance. Although increased financial input will be needed to drive improvement, the investment is likely to be highly cost-effective. Many ways of funding registries are available. A financial contribution by industry funds the successful United Kingdom National Joint Registry, and this model might be more widely applicable. Financial encouragement to hospitals from NHS governing bodies for the registration of patients may also lead to improved compliance, by helping surgeons negotiate resources. Again, this has international applicability.
There are ‘carrots and sticks’, and surgeons should note that in another similar setting regulators used punitive tactics successfully to promote the use of registries. The SWISSspine registry was introduced to monitor outcomes after the introduction of lumbar disc replacements and reported a rate of registration of 80% during the first three years. The details of patients registered by surgeons were compared with data provided by the manufacturers. Surgeons failing to submit patients to the registry risked having their certification to use lumbar disc replacements withdrawn.53 Whether regulators choose a carrot or stick approach to improve the quality of data on spine registries remains to be seen. In the absence of reliable registry data, surgeons should be aware of alternative methods of recording suspected implant-related ‘adverse incidents’, such as the Yellow Card system provided by the Medicines & Healthcare products Regulatory Agency (MHRA)54 in the United Kingdom (recommended by the BSS).
In addition to monitoring clinical outcomes of the MAGEC system, we must also learn from any failures which are encountered. The independent analysis of failed implants has proved to be valuable, particularly in the study of complex devices such as arthroplasties. Such analyses have allowed mechanisms of failure to be understood and improvements to be made to the design of components.55,56 Teoh et al7 reported the first such analysis of seven MAGEC rods explanted at their centre, identifying significant metal debris within the actuator as well as fractures of a drive pin. Although these findings are valuable, it is clear that many more rods must be analysed in greater detail before we can thoroughly understand the mechanisms of failure. Centres have been established in Newcastle- and London to analyse explanted MAGEC rods independently. We urge surgeons to ensure all explanted MAGEC rods are sent for analysis.
Surgeons using the MAGEC system have a responsibility to limit the number of failures through the careful selection of patients and close monitoring of their cases to ensure that failures are recognised. The role for the MAGEC -system in revision cases is uncertain with increased complications and limited spinal growth noted in some studies.11-13 Our audit of the BSR identified that 30% of constructs remain single rather than dual rod. This is surprising given the experience gained with traditional growing rods. In keeping with traditional growing rods, there is increasing evidence that single MAGEC rod constructs may be associated with a poorer correction of the curve and a higher rate of complications than dual rod constructs. It is apparent that dual MAGEC rod constructs should be used where possible.10,12,17,57,58
Fractures of a drive pin can be easily recognised on plain radiographs.8 Surgeons should review their cases to ensure that all patients with a fractured drive pin are identified as failed pistoning rods and are likely to be associated with significant metallosis. Though ultrasound is a useful method of measuring distraction, we would agree with other authors who suggest that plain radiographs should be obtained every six months for surveillance, as well as to record if rods fail to distract on two successive -occasions.8,59
The innovative MAGEC system is potentially a great advance in the management of patients with early-onset scoliosis. It is used widely but infrequently in most spine centres in the United Kingdom. The long-term clinical outcome is not known. Despite a mandate from NHS England to register all cases on the BSR, compliance is inadequate and does not allow a robust evaluation of the system. Notwithstanding this, experience gained in the United Kingdom suggests that failures are not uncommon and may be associated with significant metallosis. Surgeons are urged to exercise caution when selecting patients for treatment using the MAGEC system, to enter their patients prospectively on a registry, to monitor them closely and to send all explanted MAGEC rods for independent analysis.
Take home message:
- Current knowledge of the MAGEC system suggests caution in its use and close monitoring.
- Surgeons should ensure all cases are placed on a registry and send all explanted MAGEC rods for independent analysis.
1 Skaggs DL , AkbarniaBA, FlynnJM, et al.Chest Wall and Spine Deformity Study Group; Growing Spine Study Group; Pediatric Orthopaedic Society of North America; Scoliosis Research Society Growing Spine Study Committee. A classification of growth friendly spine implants. J Pediatr Orthop2014;34:260–274. Google Scholar
2 Sankar WN , AcevedoDC, SkaggsDL. Comparison of complications among growing spinal implants. Spine (Phila Pa 1976)2010;35:2091–2096. Google Scholar
3 Bess S , AkbarniaBA, ThompsonGH, et al.Complications of growing-rod treatment for early-onset scoliosis: analysis of one hundred and forty patients. J Bone Joint Surg [Am]2010;92-A:2533–2543. Google Scholar
4 Akbarnia BA , MarksDS, Boachie-AdjeiO, ThompsonAG, AsherMA. Dual growing rod technique for the treatment of progressive early-onset scoliosis: a multicenter study. Spine (Phila Pa 1976)2005;30:S46–S57. Google Scholar
5 Caldas JC , Pais-RibeiroJL, CarneiroSR. General anesthesia, surgery and hospitalization in children and their effects upon cognitive, academic, emotional and sociobehavioral development - a review. Paediatr Anaesth2004;14:910–915. Google Scholar
6 No authors listed. National Institute for Health and Care Excellence. The MAGEC System for spinal lengthening in children with scoliosis. Medical Technological -Guidance MTG18. https://www.nice.org.uk/guidance/mtg18 (date last accessed 13 March 2017). Google Scholar
7 Teoh KH , von RuhlandC, EvansSL, et al.Metallosis following implantation of magnetically controlled growing rods in the treatment of scoliosis: a case series. Bone Joint J2016;98-B:1662–1667. Google Scholar
8 Jones CS , StokesOM, PatelSB, ClarkeAJ, HuttonM. Actuator pin fracture in magnetically controlled growing rods: two cases. Spine J2016;16:287–291. Google Scholar
9 Cheung JP , CahillP, YaszayB, AkbarniaBA, CheungKM. Special article: Update on the magnetically controlled growing rod: tips and pitfalls. J Orthop Surg (Hong Kong)2015;23:383–390. Google Scholar
10 La Rosa G , OggianoL, RuzziniL. Magnetically Controlled Growing Rods for the Management of Early-onset Scoliosis: A Preliminary Report. J Pediatr Orthop2017;37:79–85. Google Scholar
11 Hosseini P , PawelekJ, MundisGM, et al.Magnetically controlled Growing Rods for Early-onset Scoliosis: A Multicenter Study of 23 Cases With Minimum 2 years Follow-up. Spine (Phila Pa 1976)2016;41:1456–1462. Google Scholar
12 Teoh KH , WinsonDM, JamesSH, et al.Magnetic controlled growing rods for early-onset scoliosis: a 4-year follow-up. Spine J2016;16:S34–S39. Google Scholar
13 Keskinen H , HeleniusI, NnadiC, et al.Preliminary comparison of primary and conversion surgery with magnetically controlled growing rods in children with early onset scoliosis. Eur Spine J2016;25:3294–3300. Google Scholar
14 Lebon J , BataillerC, WargnyM, et al.Magnetically controlled growing rod in early onset scoliosis: a 30-case multicenter study. Eur Spine J2016 December 31. (Epub ahead of print) Google Scholar
15 Thompson W , ThakarC, RoltonDJ, Wilson-MacDonaldJ, NnadiC. The use of magnetically-controlled growing rods to treat children with early-onset scoliosis: early radiological results in 19 children. Bone Joint J2016;98-B:1240–1247. Google Scholar
16 Akbarnia BA , PawelekJB, CheungKM, et al.Growing Spine Study Group. Traditional Growing Rods Versus Magnetically Controlled Growing Rods for the Surgical Treatment of Early-Onset Scoliosis: A Case-Matched 2-Year Study. Spine Deform2014;2:493–497. Google Scholar
17 Dannawi Z , AltafF, HarshavardhanaNS, El SebaieH, NoordeenH. Early results of a remotely-operated magnetic growth rod in early-onset scoliosis. Bone Joint J2013;95-B:75–80. Google Scholar
18 Heydar AM , ŞiraziS, BezerM. Magnetic Controlled Growing Rods (MCGR) As a Treatment of Early Onset Scoliosis (EOS): Early Results With Two Patients Had been Fused. Spine (Phila Pa 1976)2016 April 06. (Epub ahead of print) Google Scholar
19 Iiharreborde B MAGEC Growing Rod Study. European Paediatric Orthopaedic Society. http://www.epos.org/commitees/scientific-committee/studies (date last accessed 13 March 2017). Google Scholar
20 Efendiyev A , YilgorC, AkbiyikF, et al.Metal Ion Release During Growth-friendly Instrumentation for Early Onset Scoliosis. Spine Deform2015;3:612–629. Google Scholar
21 Visuri T , PukkalaE, PaavolainenP, PulkkinenP, RiskaEB. Cancer risk after metal on metal and polyethylene on metal total hip arthroplasty. Clin Orthop Relat Res1996;329:S280–S289. Google Scholar
22 Daley B , DohertyAT, FairmanB, CaseCP. Wear debris from hip or knee replacements causes chromosomal damage in human cells in tissue culture. J Bone Joint Surg [Br]2004;86-B:598–606. Google Scholar
23 Rolton D , RichardsJ, NnadiC. Magnetic controlled growth rods versus conventional growing rod systems in the treatment of early onset scoliosis: a cost comparison. Eur Spine J2015;24:1457–1461. Google Scholar
24 Jenks M , CraigJ, HigginsJ, et al.The MAGEC system for spinal lengthening in children with scoliosis: A NICE Medical Technology Guidance. Appl Health Econ Health Policy2014;12:587–599. Google Scholar
25 No authors listed. The Royal College of Surgeons of England. An Investigation of the Performance of the 3M Capital Hip System. https://www.rcseng.ac.uk/library-and-publications/college-publications/docs/capital-hip-system/ (date last accessed 13 March 2017). Google Scholar
26 Malchau H , BragdonCR, MuratogluOK. The stepwise introduction of innovation into orthopedic surgery: the next level of dilemmas. J Arthroplasty2011;26:825–831. Google Scholar
27 Malchau H . Introducing new technology: a stepwise algorithm. Spine (Phila Pa 1976)2000;25:285. Google Scholar
28 Birch N . Thoracolumbar spinal treatment without screws: examples of clinical proficiency. Bone Joint J2016;98-B:3–5. Google Scholar
29 McCulloch P , CookJA, AltmanDG, HeneghanC, DienerMK. IDEAL Group. IDEAL framework for surgical innovation 1: the idea and development stages. BMJ2013;346:3012. Google Scholar
30 Cook JA , McCullochP, BlazebyJM, et al.IDEAL Group. IDEAL framework for surgical innovation 3: randomised controlled trials in the assessment stage and evaluations in the long term study stage. BMJ2013;346:2820. Google Scholar
31 Ergina PL , BarkunJS, McCullochP, et al.IDEAL framework for surgical innovation 2: observational studies in the exploration and assessment stages. BMJ2013;346:3011. Google Scholar
32 No authors listed. Beyond Compliance. http://www.beyondcompliance.org.uk (date last accessed 24 April 2017). Google Scholar
33 Akbarnia BA , CheungK, NoordeenH, et al.Next generation of growth-sparing techniques: preliminary clinical results of a magnetically controlled growing rod in 14 patients with early-onset scoliosis. Spine (Phila Pa 1976)2013;38:665–670. Google Scholar
34 Cheung KM- C , Cheung JP-Y, SamartzisD, et al.Magnetically controlled growing rods for severe spinal curvature in young children: a prospective case series. Lancet2012;379:1967–1974. Google Scholar
35 Yoon WW , SedraF, ShahS, et al.Improvement of pulmonary function in children with early-onset scoliosis using magnetic growth rods. Spine (Phila Pa 1976)2014;39:1196–1202. Google Scholar
36 No authors listed. Food and Drug Administration. https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/PremarketSubmissions/PremarketNotification510k/ (date last accessed 24 April 2017). Google Scholar
37 Harrington PR . Treatment of scoliosis. Correction and internal fixation by spine instrumentation. J Bone Joint Surg [Am]1962;44-A:591–610. Google Scholar
38 No authors listed. Food and Drug Administration. 510(k) Summary Ellipse intramedullary limb lengthening system Premarket Notification Number: K101997. https://www.accessdata.fda.gov/cdrh_docs/pdf10/K101997.pdf (date last accessed 24 April 2017). Google Scholar
39 No authors listed. Food and Drug Administration. 510(k) Summary MAGEC Spinal Bracing and Distraction System Premarket Notification Number: K140178. http://www.accessdata.fda.gov/cdrh_docs/pdf14/k140178.pdf (date last accessed 13 March 2017). Google Scholar
40 Howie DW , MiddletonRG, CostiK. Loosening of matt and polished cemented femoral stems. J Bone Joint Surg [Br]1998;80-B:573–576. Google Scholar
41 Malchau H , GravesSE, PorterM, HarrisWH, TroelsenA. The next critical role of orthopedic registries. Acta Orthop2015;86:3–4. Google Scholar
42 Havelin LI , EspehaugB, VollsetSE, EngesaeterLB. The effect of the type of cement on early revision of Charnley total hip prostheses. A review of eight thousand five hundred and seventy-nine primary arthroplasties from the Norwegian Arthroplasty Register. J Bone Joint Surg [Am]1995;77-A:1543–1550. Google Scholar
43 Ellams D , ForsythO, MistryA, et al.2017). Google Scholar
44 Graves SE , DavidsonD, de SteigerR, et al.2017). Google Scholar
45 No authors listed. Center for Devices and Radiological Health. U.S. Food and Drug Administration. Strengthening our national system for medical device postmarket surveillance: Update and next steps. http://www.fda.gov/downloads/MedicalDevices/Safety/CDRHPostmarketSurveillance/UCM348845.pdf (date last accessed 13 March 2017). Google Scholar
46 Hutton M , BreakwellLMBritish Spine Registry; End of year annual report (March 2015-March 2016). http://www.britishspineregistry.com/news/ (date last accessed 13 March 2017). Google Scholar
47 No authors listed. NHS England Clinical Reference Group for Complex Spinal Surgery. Clinical Commissioning Policy: Non-Invasively Lengthened Spinal Rods for Scoliosis. https://www.england.nhs.uk/commissioning/wp-content/uploads/sites/12/2013/05/Non-Invasively-Lengthened-Spinal-Rods-for-Scoliosis.pdf (date last accessed 13 March 2017). Google Scholar
48 Kolling C , SimmenBR, LabekG, GoldhahnJ. Key factors for a successful National Arthroplasty Register. J Bone Joint Surg [Br]2007;89-B:1567–1573. Google Scholar
49 McGirt MJ , ParkerSL, AsherAL, et al.Role of prospective registries in defining the value and effectiveness of spine care. Spine (Phila Pa 1976)2014;39:S117–S128. Google Scholar
50 Fisher C , HallH, LoveL, et al.2017). Google Scholar
51 Fritzell P , StrömqvistB, HäggO. A practical approach to spine registers in Europe: the Swedish experience. Eur Spine J2006;15:S57–S63. Google Scholar
52 Roysam GS , HillA, JagonaseL, et al.Two years following implementation of the Brisish Spinal Registry (BSR) in a District General Hospital (DGH): perils, problems and PROMS. Spine J2016;16:S79. Google Scholar
53 Schluessmann E , DielP, AghayevE, et al.SWISSspine Registry Group. SWISSspine: a nationwide registry for health technology assessment of lumbar disc prostheses. Eur Spine J2009;18:851–861. Google Scholar
54 No authors listed. Yellow card. https://yellowcard.mhra.gov.uk/ (date last accessed 25 April 2017). Google Scholar
55 Langton DJ , JamesonSS, JoyceTJ, et al.AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: A consequence of excess wear. J Bone Joint Surg [Br]2010;92-B:38–46. Google Scholar
56 Langton DJ , SidaginamaleR, LordJK, NargolAV, JoyceTJ. Taper junction failure in large-diameter metal-on-metal bearings. Bone Joint Res2012;1:56–63. Google Scholar
57 Thompson GH , AkbarniaBA, KostialP, et al.Comparison of single and dual growing rod techniques followed through definitive surgery: a preliminary study. Spine (Phila Pa 1976)2005;30:2039–2044. Google Scholar
58 Yang JS , SponsellerPD, ThompsonGH, et al.Growing Spine Study Group. Growing rod fractures: risk factors and opportunities for prevention. Spine (Phila Pa 1976)2011;36:1639–1644. Google Scholar
59 Stokes OM , O’DonovanEJ, SamartzisD, et al.Reducing radiation exposure in early-onset scoliosis surgery patients: novel use of ultrasonography to measure lengthening in magnetically-controlled growing rods. Spine J2014;14:2397–2404. Google Scholar